Resistance: Rp, Rs
Range
Resolution
DCR
100/120 Hz
1 kHz
10 kHz
100 kHz
20 Ω
0.01 Ω
-
0.7 % + 8 d
-
-
-
0.001 Ω
-
-
0.7 % + 8 d 0.7 % + 8 d 0.7 % + 8 d
200 Ω
0.01 Ω
0.3 % + 3 d 0.3 % + 3 d 0.8 % + 3 d 0.3 % + 3 d 0.3 % + 3 d
2 kΩ
0.0001 kΩ
0.3 % + 3 d 0.3 % + 3 d 0.3 % + 3 d 0.3 % + 3 d 0.3 % + 3 d
20 kΩ
0.001 kΩ
0.3 % + 3 d 0.3 % + 3 d 0.3 % + 3 d 0.3 % + 3 d 0.5 % + 5 d
200 kΩ
0.01 kΩ
0.5 % + 5 d 0.3 % + 3 d 0.3 % + 3 d 0.3 % + 3 d 0.9 % + 10 d
2 MΩ
0.0001 MΩ 0.5 % + 5 d 0.5 % + 5 d 0.5 % + 5 d 2.0 % + 10 d
-
20 MΩ
0.001 MΩ
2.0 % + 8 d 2.0 % + 10 d 3.0 % + 20 d
-
-
0.01 MΩ
-
-
-
5.0 % + 80 d
-
200 MΩ
0.1 MΩ
5.0 % + 80 d 5.0 % + 80 d
-
-
-
Capacitance: Cp, Cs (When D < 0.1)
Range Resolution 100/120 Hz
1 kHz
10 kHz
100 kHz
200 pF
0.01 pF
-
-
0.8 % + 10 d 2.0 % + 10 d
2000 pF
1 pF
0.5 % + 5 d
-
-
-
0.1 pF
-
0.5 % + 5 d
0.5 % + 5 d 0.7 % + 10 d
20 nF
0.001 nF 0.5 % + 5 d
0.3 % + 3 d
0.5 % + 5 d
0.3 % + 3 d
200 nF
0.01 nF
0.3 % + 3 d
0.5 % + 5 d
0.5 % + 5 d 0.7 % + 10 d
2000 nF
0.1 nF
0.3 % + 3 d
0.3 % + 3 d
0.5 % + 5 d 3.5 % + 10 d
20 μF
0.001 μF 0.3 % + 3 d
0.3 % + 3 d 1.0 % + 10 d
-
200 μF
0.01 μF
0.3 % + 3 d 0.8 % + 10 d
-
-
0.1 μF
-
-
3.5 % + 10 d
-
2000 μF
0.1 μF
2.0 % + 10 d
-
-
-
1 μF
-
1.5 % + 10 d
-
-
20 mF
0.01 mF 1.5 % + 10 d
-
-
-
• Accuracy when D
≧
0.1: Ae×
Ae means an accuracy of rdg in the main display.
― 38 ―
― 37 ―
Temperature
coefficient
[0.15 x (specified accuracy)]/
℃
0
~
18
℃
, 28
~
50
℃
Operating
temperature/
humidity
0
~
50
℃
< 80 % RH
Storage
temperature/
humidity
-20
~
60
℃
< 80 % RH
Safety
standard
IEC61010-031
EMC
IEC61326-1
Under a condition field strength CS 3 V (0.15-80 MHz):
The following shows a total accuracy of the DC
resistance measurement.
Total accuracy = specified accuracy ± 150 dgt
In other ranges or a condition which is worse than
the above-described condition, the accuracy is not
specified.
Power source
Single alkaline 9 V battery 6LR61(IEC6LF22,
NEDA1604A) or an external AC/DC adopter
(separately available: AD-30-2)
Auto Power
Off
5 minutes after the last operation
Power
consumption Approx. 110 mW
Battery life Approx. 35 hours (Auto LCR Mode, 1 kHz)
Dimensions
without holster Approx. L 175 mm × W 80 mm × H 40 mm
with holster
Approx. L 184 mm × W 87 mm × H 45 mm
Mass
without holster Approx. 320 g
with holster
Approx. 400 g
Accessories
Clipping leads (CL-700),
Holster (H-701)
Shorting plate for the directly connectable measuring terminals
Instruction manual
Separately
available
accessories
AC/DC adopter (AD-30-2),
SMD clipping leads (CL-700SMD)
LCR USB communication unit (LCR-USB)
(PC communication sftware bundled)
8-2 Measuring Range and Accuracy
Accuracy: ±(% rdg + dgt)
" % + d " is an abbreviation.
rdg: reading
dgt: least significant digit
Temperature: 23
℃
±5
℃
Humidity:
≦
80 % R.H.
(
≦
60 % R.H. for the range of 2 MΩ, 20 MΩ, and 200 MΩ)
Accuracies after the OPEN/SHORT calibration are shown below.
" - ": the accuracy is not guaranteed.