4. Connect the test leads to make a
short circuit.
5. Press the
CAL
button again.
The display shows
[Srt]
.
6. Press the
CAL
button again.
The meter shows a countdown on the
display while it performs the SHORT
calibration. It takes 30 seconds to
complete.
After the countdown is complete,
the display should show
[PASS]
.
If the display shows [FAIL], this
procedure has to be performed again.
7. Press the
CAL
button once again to exit the OPEN/SHORT
calibration mode.
For measaurement using the SMD clipping leads "CL-700SMD"
(separately available accessory), the same procedure as above can
be taken to perform the OPEN/SHORT calibration.
[Measurement using the clipping leads "CL-700"]
1. Make sure the leads are completely
disconnected.
2. Press the
CAL
button for 2 seconds.
The display shows
[OPEn]
.
3. Press the
CAL
button again.
The meter shows a countdown on the display
while it performs the OPEN calibration.
It takes 30 seconds to complete.
After the countdown is complete, the display
should show
[PASS]
.