![Salutron HardyTest HARTIP 1800 Operation Manual Download Page 23](http://html1.mh-extra.com/html/salutron/hardytest-hartip-1800/hardytest-hartip-1800_operation-manual_3628958023.webp)
21
Figure 6-33 Calibration
Pressing button
▼
, cursor will move between 7 6 0 and Back, the standard value can be
changed from 0 to 9 by pressing
▲
button. After setting standard block value, press
►
button to enter next step.
Moving cursor to Back and press
►
button, the current operation will be given up.
You are asked to take 3 measurements on standard test block in total. After 3
measurements, press
►
button to finish the calibration. You also can give up this
calibration by pressing
▼
button.
Note: The user calibration procedure should be done every half year, if you don’t use
the tester for long time, before start to use it, you should also do a calibration.
Calibration for impact device D and DL should be operated separately. Before
calibration, you should choose impact device from parameters menu first, and then
perform calibration for this impact device.
6.8.10.
Brightness Adjustment
Choose
Brightness
from the function menu and press
►
button to enter Brightness
adjustment window.
Figure 6-34 Brightness Adjustment
Press
▼
or
▲
button to adjust brightness, and then press
►
button to return the function
menu.
6.8.11.
Factory default
Choose
Factory default
from the function menu and press
►
button to enter factory
default menu. The instrument will give the prompt “Factory default, Sure to set?” Press
▼
button to return the function menu and press
►
button to confirm to restore factory
settings. The instrument will give the prompt “Please wait”, “Successfully”. The screen
will return to the Function menu in 1 sec.
Default setup:
Hardness scale: HLD
Impact direction: down 90
o