Salutron HardyTest HARTIP 1800 Operation Manual Download Page 18

 

16 

 

6.6.2.

 

Data recall 

Choose 

Data  recall

  from  Memory  menu  and  press 

  button  to  enter  the  block  no. 

selection window. All blocks stored in the instrument are showed on the window; at most 

9  block  numbers  are  showed  on  each  screen.  “Change  Page”  is  showed  on  upper  left 

corner. Displayed content can be changed by pressing 

 or 

 button. When the page of 

block to be viewed is chosen, press 

 button to change “Change Page” showed on upper 

left corner into “Back”. Press 

 button to change the horizontal position of cursor and 

choose the block number to be viewed. Press 

 button to enter the Data recall window. 

The window includes four-row three-column. The first row is used to show block number, 

total pages and No. of current page. Measured values are stored from the second row to 

the fourth row. Press 

 or 

 button to turn page forwards or backwards to view stored 

measured  values.  At  the  same  time,  press 

  button  and 

  button  to  show  measured 

value, hardness scale, material and impact direction alternately. Press the button once to 

switch the content once and allow cycling switch.   

On the Data recall window, press 

 button to return the Block no. selection window. Press 

Back

 to return to the Memory menu. Also at the same time press 

 button and 

 button 

to return the Memory menu directly. 

 

 

 

Figure 6-21 Data Recall 

6.6.3.

 

Delete block 

Choose 

Delete  block

  from  Memory  menu  and  press 

  button  to  enter  the  block  no. 

selection  window.  All  blocks  stored  in  the  instrument  are  showed  on  the  interface.  At 

most 9 block numbers are showed on each screen. “Change Page” is showed on upper left 

corner. Displayed content can be changed by pressing 

 or 

 button. When the page of 

block to be deleted is chosen, press 

 button to change “Change Page” showed on upper 

left corner into “Back”. Press 

 button to change the vertical position of cursor and press 

  button  to  change  the  horizontal  position  of  cursor.  Choose  the  block  number  to  be 

deleted  and  press 

  button.  Confirmation  screen  will  be  showed  on  the  instrument  to 

remind if this block is deleted. Confirm to delete this block and press 

 button to delete it. 

After  the  operation  is  successful,  “Successfully”  is  showed  on  the  instrument.  Press 

 

button  to  return  the  Block  no.  selection  window.  Press 

Back

  to  return  to  the  Memory 

menu. Also at the same time press 

 and 

 button to return the Memory menu directly. 

After  the  block  is  deleted,  all  data  of  this  block  is  lost.  Therefore,  be  cautious  to  this 
operation.  After  deleting  this  block,  its  storage  space  can  be  set  as  current  block  for 

continuous service.     

6.6.4.

 

Memory clear 

Choose 

Memory clear

 from Memory menu and press 

 button, a confirmation screen 

will be showed on the  instrument.  Press 

 button to confirm and empty the memory. 

While  empty  the  memory,  the  instrument  will  give  the  prompt  “Please  wait”.  After  the 

memory is emptied, the instrument will give the prompt “Successfully”. Press

 button to 

return the Memory menu. 
After  the  memory  is  emptied,  all  blocks  and  data  stored  in  the  instrument  will  lose. 

Therefore, be cautious to this operation. 

Summary of Contents for HardyTest HARTIP 1800

Page 1: ...HARDNESS TESTER HardyTest HARTIP 1800 OPERATION MANUAL...

Page 2: ......

Page 3: ...SAMPLE OF CURVING SURFACE 7 5 6 SUPPORTING THE SAMPLES DURING TESTING 7 5 7 SAMPLES WITH CURVED SURFACES 7 6 OPERATION 8 6 1 BUTTON DESCRIPTION 8 6 2 DIAGRAM OF OPERATION 9 6 3 POWER ON THE INSTRUMEN...

Page 4: ...ional to the velocities of the magnet Leeb hardness is expressed by the following formula 1000 Vi Vr HL Where HL is Leeb Hardness Vr is the rebound velocity of the impact body Vi is the impact velocit...

Page 5: ...l probes With this optional accessory you can take measurement at very narrow surface such as slot bottom gear tooth that probe D cannot match The 3 7V Li ion rechargeable battery inside the tester ca...

Page 6: ...Recalibration allowed by user 2 4 Applications Hardness tests on installed machines or steel structures e g on heavy and large work piece or on permanently installed system parts Rapid testing of mult...

Page 7: ...MATERIALS H R C H R B H B H V H S HRA b N mm2 STEEL CAST STEEL 20 0 67 9 59 6 99 5 80 647 80 940 32 5 99 5 30 88 375 1710 ALLOY TOOL STEEL 20 5 67 1 80 898 1170 2639 STAINLESS STEEL 19 6 62 4 46 5 101...

Page 8: ...direction of impact should be perpendicular to the coupling surface For the coupling operation the following prerequisites must be fulfilled The contact surface of the sample and the surface of the b...

Page 9: ...be checked in the following manner a Medium weight samples and also heavier samples with protruding parts or thin walls should be placed on a solid support in such a manner that they do not move or fl...

Page 10: ...start to print a new page while Bluetooth printing is on Delete current reading after set up statistics Upward button Move cursor upward Change value or digit circulative from 0 to 9 Shift normal or u...

Page 11: ...9 6 2 Diagram of Operation Figure 6 2 Diagram of Operation...

Page 12: ...s required press the confirmation button on the keyboard to enter main menu Figure 6 4 Main Menu After entering main menu press or button to move among the different options press button to enter conf...

Page 13: ...t up Impact device to DL while you are using probe DL Figure 6 6 Impact Device Selection After selecting impact device Successfully will be displayed on the screen then return to the Parameters menu F...

Page 14: ...Different hardness scales will show on the screen depending on different materials The default hardness scale is Leeb hardness value HLD Figure 6 10 Hardness Scale 6 4 5 Impact Direction Choose Impact...

Page 15: ...tatistics menu and press button to enter sub menu Then press or button to choose different mean times then press button to confirm the selection and return to statistics Menu The default mean time is...

Page 16: ...s button to confirm the selection and return to statistics menu The default value of upper limit alarm is 999 HLD When the users set the value of upper limit alarm the ranges of hardness values corres...

Page 17: ...t as any block of existing blocks Press button to move the cursor and choose the value of modified group No cycling from hundreds tens units to return in turn When the cursor is moved to the option of...

Page 18: ...window All blocks stored in the instrument are showed on the interface At most 9 block numbers are showed on each screen Change Page is showed on upper left corner Displayed content can be changed by...

Page 19: ...irm Figure 6 24 Communicating by USB RS232 Figure 6 25 Communicating by Bluetooth Via USB RS232 or Bluetooth interface testing values can be downloaded to PC or printed by micro printer with our speci...

Page 20: ...nsferring on line Start software on PC and operate the relative tool to download the data from memory Note Please make sure the connection between PC and tester is successful See 11 3 6 7 4 Print bloc...

Page 21: ...s hardness scale Under Bluetooth communication mode pressing button In order to meet the measurement of requiring different direction it is also designed to change display direction which function can...

Page 22: ...apacity and storage capacity are showed on the screen After viewing them press button to return to the function menu Figure 6 32 System info 6 8 8 User info Choose User info from the function menu and...

Page 23: ...vice D and DL should be operated separately Before calibration you should choose impact device from parameters menu first and then perform calibration for this impact device 6 8 10 Brightness Adjustme...

Page 24: ...e measurement at very narrow surface such as slot bottom gear tooth that probe D cannot match Please refer to the following steps to change impact body Unscrew the support ring from the tester and let...

Page 25: ...oper way of holding is important for obtaining better readings Attention the tester must be placed against object surface firmly and perpendicularly A slight gap between support ring of tester and sur...

Page 26: ...some time Please make sure to obtain additional or replacing battery from SALUTRON Otherwise it may cause the instrument to get un accuracy value In order to avoid damage of battery due to over discha...

Page 27: ...mode 9 3 System reset The instrument doesn t response possibly due to instable power voltage or others the instrument can be reset when it does not response Take the impact body out of the instrument...

Page 28: ...ndrical HZ 11 13 20 18 5mm R 11mm 13mm R 11mm not possible R 30mm D6a HZ 12 5 17 20 20 5mm R 12 5mm 17mm HZ 16 5 30 20 20 5mm R 16 5mm 30mm spherical K 10 15 20 7 7mm R 10mm 13mm R 10mm not possible R...

Page 29: ...wer indicator green light keep bright when the battery power 50 30 three color power indicator the yellow light keep bright when the battery is less than 30 three color power indicator the red light k...

Page 30: ...s the Drive Letter of the application CD x cp2102 Drivers CP210x USB Composite Device x cp2102 Drivers CP210x USB to UART Bridge Controller Install these two drivers 11 2 Installation of Software To r...

Page 31: ...nstalled during installation install the file NET35_SP1_cn exe that is located in X dotnet_35 folder in the CD You can also download it from the Microsoft website 11 3 Operation of PC Software Double...

Page 32: ...which is in the down left corner of the software window After checking an orange check displays that means the connection is created successfully Before Checking Device After Checking Device connecte...

Page 33: ...saved 11 3 4 Test data Click Test data tab in the working area or click button from the tool bar to load the Test data working area After connection is established you can return to the testing mode...

Page 34: ...isplay choose the file and click Open the saved data will be displayed in the current data area File Print to print current data Select Print from the File menu or click button the data in current win...

Page 35: ...click Read button to read the current parameters from the tester select new parameters in Preinstall area and click Modify button to change the current parameters Note Operations in Parameter tab must...

Page 36: ...l value of standard block in the standard value box and press Change standard value The default value is 500 The modified data is saved in the application 2 Calibration To take measurement on the stan...

Page 37: ...ew form Click button to open saved stored data Click button to read stored data Click button to save current data into Microsoft Excel document Click button to print current data to the default printe...

Page 38: ...36 11 4 Quit the program After finish all the job you can click from the tool bar to quit the program...

Page 39: ......

Page 40: ...tf S ri a e l d u C T r r e o m n er M A e ll s e s e t 3 e 0 c h 7 n ik D 5 G 0 m 22 b 6 H Dr Got Frechen Tel 49 0 2234 9999960 Fax 49 0 2234 9999962 Email info salutron de www salutron de...

Reviews: