STB8827
Series Operation Manual Chapter 5 Execute LCR Operation and Some Examples
available units will be displayed in the soft key zone: p, n,
μ
, m and k. Press
[ENTER], this zone will be changed as 0.00050.
p) Move the cursor to
FREQ
65
. The following soft keys will be displayed: ON, OFF,
MEAS OPEN, MEAS SHORT and MEAS LOAD.
q) Hold the test fixture be in open status and keep user’s hands or other
interference source be far away from the test fixture. Press the soft key MEAS
OPEN to execute open correction.
r) Insert the short plate (TH26010) into the test fixture. Please ensure that the
short plate and the reeds of the test fixture have good contact.
s) Press the soft key MEAS SHORT to execute short correction.
t) Insert a standard capacitance into the test fixture. Please ensure that the pins
of the standard capacitance have good connection with the reeds of the test
fixture.
u) Press the soft key MEAS LOAD to execute load correction.
2) Note
a) Because of the different software editions, the soft keys and status
information may be different form this book, but it may not affect users’
understanding.
b) The load correction is only valid for the components with the same
specification. If the specification is changed, it is required to redo load
correction.