STB8827
Series Operation Manual Chapter 7 Command Reference
8.2.4
125
Operation
Before you use the HANDLER function, you should ensure that your instrument
has install the HANDLER interface board.The following procedure will show you
how to use the interface comparison and the list sweep comparison.
The procedure for comparison setting :
(1) Press the softkey [LIMIT TABLE]. LIMIT TABLE SETUP page will be displayed.
(2) Set the standard value and the high limit and the low limit in LIMIT TABLE
SETUP.
Refer to the LCR menu key description to see more details.
(3) Move the cursor to COMP field. The softkey area will display :
ON
OFF
(4) Press the softkey ON to enable the comparison function.
(5) Press [LCRZ], and then press the sotfkey BIN NO. or BIN COUNT, the DUT will
be tested; at the same time, you can set the counter for the DUT and the
auxiliary.
Note: COMP ON/OFF can be set in BIN COUNT page.
The procedure for list sweep comparison:
(1)Press LIST SETUP, LIST SWEEP SETUP page will be displayed.
(2)Set the sweep mode, sweep frequency point, reference value, the high limit
and the low limit. Refer to the LCR menu key description to see more details.
(3)Press [LCRZ].and then press the softkey LIST SWEEP, LIST SWEEP DISP
page will be displayed. Refer to the LCR menu key description to see more details.
Note: the following methods can be used to improve the speed:
(1)Set the range to the maximum that the capacitance may be, and lock this
range.
(2)Set Vm: OFF and Im: OFF in the MEAS SETUP page.
(3)Test the DUT in the BIN COUNT page.