38
the fiber will be shifted relative to the discharge center, ARC calibration will be
needed.
8.3 Electrode stabilize
When the environment changes dramatically, the discharge strength will become
unstable which will increase the splicing loss, especially when it changes from low
altitude to high altitude, it needs some time to stabilize the discharge strength. Under
this condition, electrode stabilize would need to be performed for several times until
it shows [stabilize finished].
Operations
Choose [electrode stabilize] in [system maintenance].
Put the prepared fiber into fusion splicer.
Press [set], it will starts to stabilize electrode automatically according to
following procedures.
a. Discharge repeated for 5 times to ensure the place of electrode.