Evoke
®
SCS System User Manual
15.1.2 Evoke eCLS (includes Case and Lead Adapters)
Table 15-2: Evoke eCLS.
Materials
eCLS body and case
ABS Plastic
Case seal
TPE
Dimensions
100 mm x 85 mm x 20 mm (3.9 in x 3.4 in x 0.8 in)
Weight
96 g (3.4 oz.)
Electrodes
24
Port 1: electrodes 1-12, Port 2: electrodes 13-24
Functional
All other functional specifications are the same as the CLS (See
specifications
Section 15.1.1)
Radio frequency
MICS band
402.45-405.55 MHz
communication
8 channels*
Centre frequency (MHz): 402.45, 402.75,
403.05, 403.35, 403.65, 403.95, 404.25, 404.55
Modulation type
Frequency Shift Keying (FSK)
Range
1.0 m (3.3 ft.)
Effective Radiated
Power (ERP)
25 µW (-16 dBm) maximum
*Channels are automatically selected when the communication session begins.
Battery
800 mAh 3.6 V Li-ion rechargeable battery
Charging
Li-Ion battery charger
Ingress Protection
IP22 Rating for protection against access of solid objects greater than
eCLS in case
or equal to 12.5 mm, and for vertically dripping water when the
device is tilted 15 degrees.
eCLS
IP30 Rating for protection against solid objects greater than or equal
to 2.5mm, and no protection against water.
IEC 60601-1 /
Type BF Applied Part
EN 60601-1
Internally Powered Medical Electrical Equipment
Classification
Continuous Operation
Storage &
Temperature:
Min: -10 °C (14 °F)
Max: 55 °C (131 °F)
Transportation
Humidity:
Min: 0% RH
Max: 90% RH
Conditions
Pressure:
70 kPa (0.69 atm)
Max: 150kPa (1.48 atm)
Operating
Temperature:
Min: 5 °C (41 °F)
Max: 40 °C (104 °F)
Conditions
Humidity:
Min: 15% RH
Max: 90% RH
Pressure:
70 kPa (0.69 atm)
Max: 106 kPa (1.05 atm)
CLIN-UMAN-005176
Revision: 4.01
Page 40 of 53