Charging Link indicator
The
‘
Charging Link
’
indicator shows the quality of the link
between the Charger coil and the CLS.
No charging
Marginal
Acceptable
Ideal
possible.
alignment.
alignment for alignment for
Charging
charging.
charging.
Charger will
possible but
beep every
Charger will
second.
may take
beep once if
longer.
the coil
alignment
falls below
this quality.
Note:
To improve the quality of the link between the Charger coil
and the Evoke CLS, move the coil closer to the implant.
Charger Battery Level indicator
The
‘
Charger Battery Level
’
indicator shows the battery level in
the Charger.
Caution:
As the Charger provides
the charge for recharging the CLS
battery, it is important to start the
charging session with a full Charger
battery.
If the charge falls to one bar, recharge
the Charger.
Note:
Always recharge the Charger at the end of each CLS
charging session to ensure you start the next charging session
with a full Charger.