19
to providing circuit diagrams, parts lists and anything else necessary to calibration and repairs the device,
only to a qualified and authorized personnel in those activities.
WARNINGS
All interventions must be recorded in Table sheet machine in order to ensure the maximum safety of the
device.
WARNINGS
This device is not intended for use by people (including children) with restricted physical, sensory or mental
skills or a lack of experience and/or a lack of knowledge, unless they are supervised by a person who has
responsibility for their safety or they receive instructions from this person on how to use the device.
WARNINGS
Please placed out of children
’
s and pets reach to prevent device damage.
NOTE
The following table contains the summary of the maintenance operations
TYPE
ACTION
AUTHORITY
FREQUENCY
Preventive
Visual inspection of the integrity of
the charging cable (if there is
deterioration contact the technical
support service)
Operator
Weekly
Preventive
Cleaning the inside, verification of
functionality, review and calibration
Authorized
laboratory
Annually
Preventive
Instrumental tests of currents
leakage and electrical safety
Authorized
laboratory
Annually
Preventive
Batteries replacemement
Authorized
laboratory
At 300 charge
cycles or after 24
months use
Ordinary
Cleaning of sensor lens
Operator
Each use
Ordinary
Visual inspection of the integrity of
patient electrodes and sensor lens
Operator
Each use
Extraordinary
Disposing accessories and / or
obsolete device in accordance with
local regulations
Operator
When necessary
WARNINGS
Tampering with the device by the user and the failure of the maintenance table above involve the
termination of the guarantee.