18
9
ORDINARY AND EXTRAORDINARY PREVENTIVE MAINTENANCE
CLEAN AND CARE
•
Regularly clean your strap and wrist—especially after working out or sweating. Rinse the strap with
water. Do NOT use hand soap, body soap, dish soap, hand sanitizers, cleaning wipes or household
cleaners, which could get trapped beneath the band and irritate skin. Instead use a soap-free
cleanser to clean the wristband. Always dry the band well before putting it back on.
•
For tough spots, stains or buildup on your strap, scrub with a wet, soft-bristled toothbrush.
•
Prolonged rubbing and pressure may irritate the skin, so give your wrist a break by removing the
strap for an hour after extended wear.
Other Tips
•
If you have eczema, allergies or asthma, you may be more likely to experience a skin irritation or
allergy from this device.
•
Whether you have the conditions above or not, if you start to experience redness or skin irritation
on your wrist, remove your device. If symptoms persist longer than 2-3 days of not using your
device, contact a dermatologist.
•
If you sweat for more than two hours while wearing your device, be sure to wash your strap and
your wrist using the directions above to avoid skin irritation.
PREVENTIVE MAINTENANCE
: to ensure the correct operation and safety of the user and the operator, you
must subject the device to some checks and periodic inspections.
Therefore it is important to check the device by authorized staff with maximum intervals of 12 months or if
in case of any malfunction.
BATTERY REPLACEMENT:
WARNINGS
The battery must / may be replaced after 300 charge cycles or after 24 months use.
To replace the CorX battery contact the technical support Saluber MD.
ORDINARY MAINTENANCE:
the operator must check the wear of patient electrodes through the time, and
promptly replace them (by authorized dealer), if there is a need.
WARNINGS
In case of evident abnormalities, do not use the device.
EXTRAORDINARY MAINTENANCE
: the extraordinary maintenance must be realized by internal staff of
Saluber MD or qualified personnel, trained and authorized by Saluber MD. The manufacturer is committed