15
Electromagnetic Immunity
The device is intended for use in the electromagnetic environment specified below.
WARNINGS
The user must ensure that it is used in that specific environment.
Failure to observe may generate risk situations.
Immunity Test
Test Level
IEC 60601
Conformity
Electromagnetic Environment - Guidance
CEI EN 61000-4-2
Electrostatic discharge
(ESD)
+6 kV in contact
compliant
Floors should be made of wood, concrete or
ceramic. If the floors are covered with
synthetic material, the relative humidity
should be at least 30%.
+8 kV in air
compliant
CEI EN 61000-4-4
Electric fast
transient/Burst
+2 kV for the power
line
compliant
The quality of the network voltage should be
that of a typical commercial or hospital
environment.
CEI EN 61000-4-5
Extra Voltages
+1 kV in differential
mode
compliant
The quality of the network voltage should be
that of a typical commercial or hospital
environment.
(applies to the step of charging the batteries)
+2 kV in common
mode
compliant
CEI EN 61000-4-11
Voltage dips, short
interruptions and voltage
variations on power
supply input lines.
0% Ut (dip
100%)
for 0,5 cycles;
N.A.
Device powered only by battery
40% Ut (dip
60%)
for 5 cycles
N.A.
70% Ut (dip
30%)
for 25 cycles
N.A.
0% Ut (dip
100%)
for 5 seconds
N.A.
CEI EN 61000-4-8
Magnetic field with
network frequency
(50 Hz)
3 A/m
compliant
The magnetic fields should have levels
characteristic of a typical location in a typical
commercial or hospital
Note: Ut is the network voltage before the application of the test level.
The second table (rif.202 of the standard cited above) concerns the various tests of immunity to interference, highlighting the type
and levels of the test, in addition to reporting important observations and recommendations for the user. The immunity is the
ability of a device to not be disturbed by electromagnetic disturbances in the environment of use, whether they are transmitted by
cable (ducts) than in air (radiated).