CARE OF THE INSTRUMENT
7.10
Revised 10/02/2012
Retort Sight Tube Cleaning
Retort Sight Tubes should be cleaned weekly to ensure
reagent in Retorts #1 and #2 is filled and maintained at the
appropriate levels for proper processing of tissue speci-
mens, and to guard against reagent overflow.
To clean the Retort Sight Tubes:
1.
Open the Retort Access Door (Figure 7-R). The
Access Door supports will retain the door in the open
position.
Figure 7-R
2.
Remove sight tube screw (Figure 7-S)
Figure 7-S
NOTE: Open Retort #1 and #2 lids manually, as neces-
sary, to facilitate cleaning the retort sight tubes.
3.
Insert the sight tube cleaning brush (see
“Section 6,
Consumables and Accessories”
for details) into the
sight tube, and scrub up and down several times to
remove debris from the wall of the sight tube (Figure
7-T).
Figure 7-T
NOTE: Debris removed from the sight tube walls during
cleaning will accumulate in the bottom of the sight tube.
Be sure to use the retort cleaning pipette to extract and
expel any accumulated debris.
4.
Insert the retort cleaning pipette (see
“Section 6,
Consumables and Accessories”
for details) into the
trap at the bottom of the sight tube. Draw out the re-
maining reagent.
5.
Remove the pipette from the sight tube and discard
reagent.
6.
Repeat steps 4 and 5 as needed to extract and expel
all debris from the sight tube.
7.
Repeat steps 2 through 5 for Retort #2.
8.
Manually close the retort lids.
9.
Close the Retort Access Door.
Remove
this screw
Summary of Contents for Tissue-Tek Xpress x120
Page 1: ...Operating Operating Manual Manual Rapid Tissue Processor Tissue Tek Xpress x120 ...
Page 3: ......
Page 20: ...INTRODUCTION 1 14 Revised 10 02 2012 ...
Page 24: ...INSTALLATION 2 4 Revised 10 02 2012 ...
Page 54: ...RECOVERY PROCESS 5 4 Revised 10 02 2012 ...
Page 82: ...A 8 Revised 10 02 2012 ...