5. Troubleshooting
91
Cleaning the Finger Plate
Remove paraffin attached to or around the finger plate using the scraper or
cotton swabs.
Paraffin attached to or around the finger plate will prevent the plate from
operating properly.
Cleaning the Hot Plates and Cold Spot
1. Remove paraffin attached to the hot plate and cold spot, using the
scraper wrapped with a lint-free paper.
2. For paraffin or foreign matters in the grooves of the hot plate, remove it
using cotton swabs.
CAUTION:
The hot plates are very hot. Wear rubber gloves or other
protective gears to prevent burns.
Finger plate
Scraper wrapped with
a lint-free paper
Cold spot
Hot plate
Cotton swab
CAUTION