5. Troubleshooting
83
Error
code
Overview
Displayed
message
Description
Instrument operation
when the error occurs
Solution
E040
Low battery
caution
Battery voltage
has dropped
The voltage of the clock
battery dropped below
the specified value.
Operation continues.
C o n t a c t t h e S a k u r a
instrument distributor or
representative and replace
t h e b a t t e r y w i t h i n o n e
month.
E041
Clock reset
The clock was
reset.
Unable to restore time
data. Button battery is
dead.
If the instrument was
operating in the auto
m o d e , o p e r a t i o n
pauses. If the instrument
was operating in the
manual mode, operation
continues.
C o n t a c t t h e S a k u r a
instrument distributor or
representative and replace
the battery.
E042
Invalid clock
The time data is
incorrect
The clock generated an
error. The battery case
came off.
Operation continues.
Check if the date and time
displayed on the screen is
correct. If the date and time
is not correct, set the correct
date and time. If the problem
persists, contact the Sakura
instrument distributor or
representative and replace
the battery.
E050
Momentary
power failure
Momentary
power outage
A m o m e n t a r y p o w e r
failure occurred.
O p e r ation c ontinue s
f r o m t h e O p e r a t i o n
screen. If the light was
o n w h e n t h e p o w e r
was cut off, operation
continues with the light
turned off.
Check the molten state of
paraffin, and then resume
using the instrument.
E100
Abnormal
tempera-ture
sensor
Temperature
sensor error
S i g n a l s c a n n o t b e
r e c e i v e d f r o m t h e
temperature sensor.
Operation aborts.
Turn off the power to the
i n s t r u m e n t a n d c o n t a c t
t h e S a k u r a i n s t r u m e n t
distributor or representative.
E101
E102
A n a b n o r m a l s i g n a l
was received from the
temperature sensor.
Operation aborts.
Turn off the power to the
instrument. When paraffin in
the problem location starts
solidifying (below 60ºC, as
a guide), turn on the power
again. (The power should be
off for at least 10 seconds.)
If th e p r o b l e m p e r s i s t s ,
turn off the power to the
i n s t r u m e n t a n d c o n t a c t
t h e S a k u r a i n s t r u m e n t
distributor or representative.
( C h e c k t o s e e i f a n y
equipment that generates
s t r o n g e l e c t r o m a g n e t i c
waves is nearby.)
E103
Abnormally
high
tempera-ture
Abnormally high
tempera-ture
The temperature of each
loc ation exc e e ds the
allowable value.
( T h e t e m p e r a t u r e i s
highe r tha n the E010
level.)
Operation aborts.
Turn off the power to the
instrument. When paraffin in
the problem location starts
solidifying (below 60ºC, as
a guide), turn on the power
again. (The power should be
off for at least 10 seconds.)
If th e p r o b l e m p e r s i s t s ,
turn off the power to the
i n s t r u m e n t a n d c o n t a c t
t h e S a k u r a i n s t r u m e n t
distributor or representative.
( C h e c k t o s e e i f a n y
equipment that generates
s t r o n g e l e c t r o m a g n e t i c
waves is nearby.)
High temperature molten
paraffin was added.
E901
Abnormal
nonvolatile
memory
Memory error
during data
update
Data could not be saved
because the nonvolatile
m e m o r y o p e r a t e d
abnormally.
Operation aborts.
Operation aborts. Turn off
the power to the instrument
a n d c o n t a c t t h e S a k u r a
instrument distributor or
representative.
E902
E903
E904
Summary of Contents for Tissue-Tek TEC 6
Page 5: ...Table of contents 5 ...
Page 6: ...1 Introduction ...
Page 31: ...1 Introduction 31 ...
Page 32: ...2 Installation ...
Page 54: ...3 Explanation of Operations ...
Page 74: ...4 Error Log ...
Page 76: ...5 Troubleshooting ...
Page 95: ...5 Troubleshooting 95 ...
Page 96: ...6 Replacement Parts ...
Page 98: ...7 Glossary of Terms ...
Page 102: ...102 ...
Page 103: ...M01 021E 02 ...
Page 104: ...Please visit our websites www sakura finetek com www sakuraus com www sakura eu ...