![Sakura Tissue-Tek TEC 6 Operating Manua Download Page 43](http://html.mh-extra.com/html/sakura/tissue-tek-tec-6/tissue-tek-tec-6_operating-manua_801569043.webp)
2. Installation
43
Setting the Temperature
1. Turn the Embedding Module power switch on.
NOTE:
The power switch of the Cryo Module may be off. Also note that
Cryo Module temperatures can be set even when the control
cable is not connected.
2. Touch the location corresponding to the temperature you want to set.
3. Touch the " "/" " buttons to change the set temperature. Touching
the "
" button saves the setting, after which the display returns to the
Operation screen.
Touch the " " button, and the display will return to the Operation screen
without the setting being saved.
NOTE:
The current temperature displayed on the Operation screen will
blink until it falls 3/-1˚C of the set temperature.
CAUTION:
Set the temperatures of the heating parts of the instrument
so that the tissues, paraffin, cassettes and base molds can
be kept at the same, appropriate temperature as much as
possible. A large temperature difference prevents the paraffin
block from remaining in a uniform state, and the block may
crack.
CAUTION
Summary of Contents for Tissue-Tek TEC 6
Page 5: ...Table of contents 5 ...
Page 6: ...1 Introduction ...
Page 31: ...1 Introduction 31 ...
Page 32: ...2 Installation ...
Page 54: ...3 Explanation of Operations ...
Page 74: ...4 Error Log ...
Page 76: ...5 Troubleshooting ...
Page 95: ...5 Troubleshooting 95 ...
Page 96: ...6 Replacement Parts ...
Page 98: ...7 Glossary of Terms ...
Page 102: ...102 ...
Page 103: ...M01 021E 02 ...
Page 104: ...Please visit our websites www sakura finetek com www sakuraus com www sakura eu ...