This product is
medical device.
(Model name : TRAUS SIP20)
Make sure to read this user manual before using.
This user manual is to assure proper installation and use.
Pay attention to read this user manual in order to making the best use of this product
and to assure the prolonged lifespan before using.
Specially, pay attention the contents with marks such as , and .
<User>
• Qualified Professional
<Intended use>
• TRAUS SIP20 is intended for use in dental surgery, implantology, maxilla-facial surgery and
endodontics for treatment of dental hard tissue and mechanical rotating root canal preparation.
<Classification of Equipment>
• Type of protection against electric shock: ClassⅠ equipment
• Degree of protec
tion against electric shock: Type B applied part
• Classification
according to the degree of protection against ingress of water as detailed
in the current edition of IEC 60529
- Control Box: IPX0
- Foot Controller: IPX8
• Not suitable product in the atmosphere that exist flammable anesthetic mixture with air
or with oxygen.
• User sterile product (non-sterile product at shipment)
• Classification according to mode of operation: : Non-continuous operation
• Applied parts: Bur
<Use Time>
• Loading time: 2 minutes
• Resting time: 15 minutes
<Operation & Storage Condition>
INTRODUCTION
Operating Condition
Temperature (℃)
10~40
Relative Humidity (%)
10~80
Atmospheric Pressure (hPa)
700~1,060
Transport and Storage Condition
Temperature (℃)
-20~60
Relative Humidity (%)
10~90
Atmospheric Pressure (hPa)
500~1,060