■ Error and Remedy
: In case of stopping handpiece by not working & using error, each error code may be shown below.
■ Fuse replacement
Cause & Remedy
Error code
Cause : The foot controller connection is not proper or the foot controller is defective.
Remedy : Make sure that the foot controller is properly connected.
Contact the manufacturer if the remedies do not activated.
Cause : The motor connection is not proper or the motor is defective.
Remedy : Make sure that motor properly connected.
Contact the manufacturer if the remedies do not activated.
Cause : Torque is overloaded than the setting
values during the surgery.
Remedy : Press the speed control pedal to release the error.
Contact the manufacturer if the remedies do not activated.
Motor Check!
Warning
E2
Torque Limit!
Warning
E3
Foot Check!
Warning
E1
- Check the fuse if the control box doesn’t work.
- If necessary contact the manufacturer.
Fuse Box