background image

17

1. Cleaning

■ 

Angle Handpeice

  

① Rotate the head in a tap water about 10 seconds with maximum speed to remove the remains

      (blood, physiologic saline solution, etc.) and rinse.
     

 Do not put tap water in the motor.

  

 Disconnect the angle handpiece to the motor.

  

 Remove contaminants on the angle handpiece using a nylon brush after soaking for 20 minutes

       in Enzymatic Detergent.
  

 Rinse the angle handpiece using the tap water for 3 minutes to remove Enzymatic Detergent.

  

⑤ Wash the angle handpiece for 10 minutes in the ultrasonic cleaner containing Enzymatic Detergent.

  

 Rinse the angle handpiece for 3 minutes in the ultrasonic cleaner containing purified water repeatedly

       subjected 3 times. (Change the purified water each time.)

  

⑦ Wipe using dust-absorbent after the product soaked in alcohol(70%) for 30 seconds.

  ⑧ Please visually inspect the angle handpiece in a bright place after cleaning. If the angle handpiece is

      still soiled after cleaning, the cleaning process must be repeated .

■ 

Control Box, Foot Controller , Motor

  Wipe the surface of motor with alcohol swab or paper towel after removing the dirt.

  ※

 Pay attention to not enter the water into the motor.

              - Proteolytic enzyme detergents are recommended as Enzymatic Detergent.
              - It is recommended to use ethanol for alcohol (70%). 
              - If the device is determined not to be visually clean at the end of the cleaning step, the user

                should either repeat the relevant previous cleaning steps or safely dispose of the device,
                so that a visibly soiled device is not used again. 

MAINTENANCE OF THE PRODUCT

2. Sterilization

■ 

Sterilization Product List

  

 Angle handpiece

  

 Motor (Assembling with motor cap)

■ 

Sterilization Method

  ①

 Dry the applicable products completely.

  

 Put the applicable products in a sterile bag and seal tightly.

     ※

 Make sure to assemble the motor cap for autoclave on the motor.

     ※

 Do not disassemble the motor and motor cord when sterilizing.

  ③ Put the sterile bag in the sterilizer and sterilization for 4 minutes with 132℃

, and let it dry for 30 minutes.

* Maximum sterilization number is 250 times.

* Moist steam sterilization only. Otherwise the products might be damaged.
* Make sure to use the sterilization bag and seal tightly.
* Make sure to dry the applicable products before sterilization.
* Motor could be damaged in sterilization because of the moisture inflow or else, 
   so should be careful to handle it.
* Sterilizing the angle handpiece with blood or other remains may cause the breakage 
   of the angle handpiece
   Therefore, it is necessary to clean the angle handpiece before sterilization

* Irrigation tube and Y-tube are disposable. Do not re-use
* Keep dry a connector and socket of the control box.

* Before using, dry the product completely.

Sterilization

 

Temperature

 

Time

 

Dry

 

Moist heat sterilization 

1

32

 

4

 minutes 

30 minutes 

Summary of Contents for TRAUS SIP20

Page 1: ...USER MANUAL www saeshin com SIP20...

Page 2: ...NTS PART NAME OF CONTROL BOX BUTTONS ON THE OPERATION PANEL LCD WINDOW BUTTONS ON THE FOOT CONTROLLER CONNECTION OF EACH PART OPERATION PROCESS MAINTENANCE OF THE PRODUCT ATTENTION PRODUCT SPECIFICATI...

Page 3: ...otection against electric shock Class equipment Degree of protection against electric shock Type B applied part Classification according to the degree of protection against ingress of water as detaile...

Page 4: ...ste after use Contact manufacturer if an incorrect display appears on the screen Do not drop or bang the product Do not damage the irrigation tube as it will not work properly due when the tube is bro...

Page 5: ...cause electric shock and or fire Unauthorized modification and dismantling is prohibited Do not use a solvent such as thinner or benzene for cleaning Please use an absorbent or alcohol to wipe the co...

Page 6: ...oller The ergonomically designed Foot Controller with main functions can be used during any surgery for user s convenience Setting with SETUP Mode Auto Calibration Sound Volume Control and Program Res...

Page 7: ...olution is not included in a set Irrigation tube and Y Tube are disposable Do not use them again Component Component Control Box Internal Spray Nozzle Optional Angle Handpiece BLDC Motor Optional Moto...

Page 8: ...NTROL BOX Foot Controller Connector Fuse Box Main Power Switch Menu Button OK Button Control Button Gear Ratio Button Optic Button Irrigation Button Rotation Direction Button LCD Display Hanger Hole M...

Page 9: ...Optic motor Control Button This is used for selecting the program number SETUP menu and adjusting the rpm torque The setting range of rpm and torque are different from gear ratio OK Button This is us...

Page 10: ...the corresponding menu is selected the RPM can be adjusted using the and buttons 7 Torque Menu Display When the corresponding menu is selected the torque can be adjusted using the and buttons 8 SETUP...

Page 11: ...n after releasing the button 2 Operation Direction Button Use the button for selecting the operation direction Press the button change the rotary direction 3 Irrigation Pump Button Use the button for...

Page 12: ...sure the power is off Otherwise it might cause a malfunction CONNECTION OF EACH PART 2 Foot Controller Connection Insert the foot controller connector Check if the holes and pins of foot controller co...

Page 13: ...r You must close the cover and use it as your fingers may get stuck Make sure the pump roller cover is correctly close Insert the irrigation tube correctly Locate the Irrigation tube correctly when in...

Page 14: ...ry type If the pump is in operation mode but water does not come out of the handpiece the tube may be broken or damaged If the user did not remove the internal spray nozzle and pressed the button or o...

Page 15: ...e bur Insert the bur to the angle handpiece Insert certainly the bur while pushing the head button on angle handpiece Release the button Pull out the bur to check it 2 Disconnecting the bur Pull out t...

Page 16: ...go to program number menu 2 Press the Control Button to select the desired program number 3 After setting the desired value by referring to the operation method keep pressing OK Button for 3 seconds...

Page 17: ...Other Function SETUP Press the Menu Button to go to SETUP Menu and press the OK Button Press the Control Button to the desired item Press the OK Button in SETUP to save and return to main After conne...

Page 18: ...ies do not activated Cause The motor connection is not proper or the motor is defective Remedy Make sure that motor properly connected Contact the manufacturer if the remedies do not activated Cause T...

Page 19: ...is determined not to be visually clean at the end of the cleaning step the user should either repeat the relevant previous cleaning steps or safely dispose of the device so that a visibly soiled devi...

Page 20: ...e product Do not put any physical pressure on the all cords 4 When the products are not using for a long term please be advised to the followings Turn off the power and unplug the cord Check the cord...

Page 21: ...Handpiece 4 Foot Controller Model Name IPX8 TRAUS FUS10 Classification according to the degree of protection against ingress of water IEC 60529 Model Name Gear ratio Optic Non optic Bur Replacement T...

Page 22: ...the other device s are connected Consult the manufacturer or filed service technician for help Electromagnetic Emissions The TRAUS SIP20 is intended for use in the electromagnetic environment specifie...

Page 23: ...t 8 angles 0 UT for 1 cycle 70 UT for 25 30 cycles 0 UT for 250 300 cycle 8kV contact 15kV air 2kV for power supply lines 1kV for input output lines 1kV differential mode 2kV common mode 0 UT for 0 5...

Page 24: ...o AM FM radio broadcasts and TV broadcasts cannot be determined exactly by theory To assess the electromagnetic environment due to fixed radiofrequency emitters an electromagnetic environment measurem...

Page 25: ...lection from structures objects and persons 150 kHz 80 MHz d 1 2 P Separation distance according to frequency of transmitter m 800 MHz 2 7 GHz d 2 3 P 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2...

Page 26: ...use or improper use and treatment Use the product with incorrect input voltage AC voltage Drop the device during setting moving using Use the not recommended consumables or accessories Malfunction aft...

Page 27: ...ctric shock Type B Manufacturer Consult Instruction For Use General Caution Sign General Warning Sign Alternating Current Power On Main power switch Power Off Main power switch This product should not...

Page 28: ...53 580 0999 1 52 5 Tel 053 587 2345 Fax 053 580 0918 SOM S048S E Rev 3 2022 07 27 www saeshin com traus saeshin com MERIDIUS MEDICAL EUROPE LIMITED Unit 3D North Point House North point Business Park...

Reviews: