2. Sterilization

    Sterilization product list

  

● 

Angle handpiece, Prophy head

Sterilization method

  ① Dry the applicable protucts completely after cleaning.

  ② Put the product in a sterile bag and seal tightly.

  ③ Put the sterile bag in the sterilizer and sterilization for 4 minutes with 132℃, and let it dry for 30 minutes.

- This product is designed to withstand 250 times of re-sterilization processes.

- This product recommends the moist heat sterilization (autoclave), other sterile methods are not

  guaranteed the effectiveness.

- Make sure to use the sterilization bag and seal tightly.

- Make sure to dry the applicable products before sterilization.

- If sterilization is performed with blood or etc. remaining inside, it may break down. Be sure to clean and

  lubricate the angle handpiece and prophy head before sterilization.

- Before using, dry the product completely.

- Do not sterilize the motor handpiece as it is not target to sterilization.

Temperature

Sterilization

Time

Dry time

132℃

4 minutes

30 minutes

Moist heat sterilization

132

1. Cleaning

  

 Angle handpiece

    ① Rotate the head in a tap water about 10 seconds with maximum speed to remove the remains

        (blood, physiologic saline solution, etc.) and rinse.

      ※ Do not put tap water in the motor handpiece.

    ② Disconnect the angle handpiece to the motor handpiece.

    ③ Remove contaminants on the angle handpiece using a nylon brush after soaking for 20 minutes

        in Enzymatic Detergent.

    ④ Rinse the angle handpiece using the tap water for 3 minutes to remove Enzymatic Detergent.

    ⑤ Wash the angle handpiece for 10 minutes in the ultrasonic cleaner containing Enzymatic Detergent.

    ⑥ Rinse the angle handpiece for 3 minutes in the ultrasonic cleaner containing purified water repeatedly

        subjected 3 times. (Change the purified water each time.)

    ⑦ Wipe using dust-absorbent after the product soaked in alcohol(70%) for 30 seconds.

  

 Prophy head

    ① Rinse the using the tap water after disconnect the prophy head to the angle handpiece.

    ② Remove contaminants on the handpiece using a nylon brush after soaking for 20 minutes

        in Enzymatic Detergent.

    ③ Rinse the prophy head using the tap water for 3 minutes to remove Enzymatic Detergent.

    ④ Wash the prophy head for 10 minutes in the ultrasonic cleaner containing Enzymatic Detergent.

    ⑤ Rinse the prophy head for 3 minutes in the ultrasonic cleaner containing purified water repeatedly

        subjected 3 times. (Change the purified water each time.)

    ⑥ Wipe using dust-absorbent after the prophy head soaked in alcohol(70%) for 30 seconds.

※ If foreign substances around the prophy head are not easily removed, use a brush to remove it.

  

 Motor handpiece & Charger

     

Wipe the surface of motor handpiece and charger with alcohol swab or paper towel after removing the dirt.

     ※ Pay attention to not enter the water into the motor handpiece and charger.

MAINTENANCE OF THE PRODUCT

- Do not use solubilizers like thinner or benzene for cleaning.

10

Summary of Contents for Traus ENDO

Page 1: ...www saeshin com INSTRUCTION MANUAL ENDO...

Page 2: ...N PRODUCT SPECIFICATION ELECTROMAGNETIC COMPATIBILITY INFORMATION WARRANTY SYMBOL Table of contents This product is cordless handpiece for root canal treatment This motor has to be used only in hospit...

Page 3: ...pment Type of protection against electric shock Class II equipment Degree of protection against electric shock Type BF applied part Classification according to the degree of protection against ingress...

Page 4: ...tion the handpiece correctly Pushing it into the charger forcibly in the incorrect direction may cause damage Take care not to put needle pin into the charging terminal of the battery charger Place th...

Page 5: ...lace If the batteries leak do not allow the liquid to come into contact with your skin eyes or clothes If the batteries leak from the motor handpiece or there change of the exterior or part discolor o...

Page 6: ...1 10 1 16 1 20 1 32 1 The free running speed of the angle handpiece shall be in accordance with the manufacturer s instructions at a tolerance of 10 as specified Refer to ISO 14457 Prophy products mus...

Page 7: ...lied to the motor it may automatically step to prevent overheating In this case leave the motor for a while until it cools down Please check all components are included PRODUCT COMPONENTS PART NAME OF...

Page 8: ...display on the screen or 2 Battery display Shows 3 steps for charging amount 3 Beep display Beep sound function each operating condition 4 Program display Shows the program number currently selected...

Page 9: ...ile inwards by matching the home position until the angle handpiece clicks Disconnection Press the button on the head to pull out the file Turn off the power when insert or remove the file Pull out th...

Page 10: ...eep sounds 4 Plug the motor handpiece into the charger with it facing forward The charging lamp lights up and charging starts 5 Orange light is on during charging and green light is on when fully char...

Page 11: ...act dealer The temperature of motor handpiece is high The temperature of battery is high The charger does not work The charge lamp does not light The plug of the charger is not inserted into the outle...

Page 12: ...using a nylon brush after soaking for 20 minutes in Enzymatic Detergent Rinse the angle handpiece using the tap water for 3 minutes to remove Enzymatic Detergent Wash the angle handpiece for 10 minut...

Page 13: ...n lubrication please hold the angle handpiece firmly because the angle handpiece can go out by the spray pressure Do not insert the oil into the motor handpiece It may cause failure When cleaning hand...

Page 14: ...2 4V Approx 2 5 hours Approx 9 hours No load PRODUCT SPECIFICATION Model name Input Output Power consumption TRAUS XEW10 AC100 240V 50 60HZ 0 1A DC 4 5V 7 5VA Model name Gear ratio Bur replacement ty...

Page 15: ...he other device s are connected Consult the manufacturer or filed service technician for help Electromagnetic Emissions The TRAUS ENDO is intended for use in the electromagnetic environment specified...

Page 16: ...t 8 angles 0 UT for 1 cycle 70 UT for 25 30 cycles 0 UT for 250 300 cycle 8kV contact 15kV air 2kV for power supply lines 1kV for input output lines 1kV differential mode 2kV common mode 0 UT for 0 5...

Page 17: ...FM radio broadcasts and TV broadcasts cannot be determined exactly by theory To assess the electromagnetic environment due to fixed radiofrequency emitters an electromagnetic environment measurement...

Page 18: ...1 At 80 MHz and 800 MHz the higher frequency range applies Note 2 These specifications may not be applicable in all situations Electromagnetic propagation is affected by absorption and reflection fro...

Page 19: ...voltage Drop the device during setting moving using Use the not recommended consumables or accessories Malfunction after repairing the product at the not recommended repair shop Normal wear of the con...

Page 20: ...580 0918 Tel 82 53 587 2341 Fax 82 53 580 0999 www saeshin com traus saeshin com This product is medical device MERIDIUS MEDICAL EUROPE LIMITED Unit 3D North Point House North point Business Park New...

Reviews: