- Check if the angle handpiece/prophy head connected properly to the motor handpiece.

CONNECTION OF EACH PART 

1. Connection of charger

① Connector the power cord to the power cord connection socket on the back of the charger. (

Picture

 1)

② Insert the power cord and plug in.

③ Plug the motor handpiece into the charger with it facing forward. (

Picture

 2)

2. Connection of angle handpiece/prophy head

  Angle handpiece/prophy head can be connected to motor handpiece at any position.

  Align angle handpiece/prophy head positioning pins with the positioning slots of motor handpiece and

  insert the head until “Click” pull it out axially for removing.

  When removing, hold the motor handpiece and pull the angle handpiece/prophy head.

3. Connection and disconnection with file

① Connection : Slide the file inwards by matching the home position until the angle handpiece clicks. 

② Disconnection : Press the button on the head to pull out the file.

- Turn off the power when insert or remove the file.

- Pull out the file to confirm locking it after connecting the file.

- Clean the chuck of angle handpiece at all times. The file can lose its balance or chucking power can

  be weakened if there is rust inside of the chuck.

① Remove the rubber cap from the product.

② Use a cross screwdriver to loosen the screw in that portion.

③ Open the battery cover to remove the battery. Check the positive and negative poles of the product and install

    new batteries properly. Reassemble the battery cover in reverse order.

Prophy head

Angle handpiece

Prophy head

Angle handpiece

(

Picture

 1)

(

Picture

 2)

DPA bur

Prophy head

4. Battery change

- Use batteries of the correct rating. (DC 1.2V, 750mAh / 2 pcs)

- When replacing the battery, make sure to check the direction of the + poles and – poles.

Rubber cap

Battery cover

Screw

7

Summary of Contents for Traus ENDO

Page 1: ...www saeshin com INSTRUCTION MANUAL ENDO...

Page 2: ...N PRODUCT SPECIFICATION ELECTROMAGNETIC COMPATIBILITY INFORMATION WARRANTY SYMBOL Table of contents This product is cordless handpiece for root canal treatment This motor has to be used only in hospit...

Page 3: ...pment Type of protection against electric shock Class II equipment Degree of protection against electric shock Type BF applied part Classification according to the degree of protection against ingress...

Page 4: ...tion the handpiece correctly Pushing it into the charger forcibly in the incorrect direction may cause damage Take care not to put needle pin into the charging terminal of the battery charger Place th...

Page 5: ...lace If the batteries leak do not allow the liquid to come into contact with your skin eyes or clothes If the batteries leak from the motor handpiece or there change of the exterior or part discolor o...

Page 6: ...1 10 1 16 1 20 1 32 1 The free running speed of the angle handpiece shall be in accordance with the manufacturer s instructions at a tolerance of 10 as specified Refer to ISO 14457 Prophy products mus...

Page 7: ...lied to the motor it may automatically step to prevent overheating In this case leave the motor for a while until it cools down Please check all components are included PRODUCT COMPONENTS PART NAME OF...

Page 8: ...display on the screen or 2 Battery display Shows 3 steps for charging amount 3 Beep display Beep sound function each operating condition 4 Program display Shows the program number currently selected...

Page 9: ...ile inwards by matching the home position until the angle handpiece clicks Disconnection Press the button on the head to pull out the file Turn off the power when insert or remove the file Pull out th...

Page 10: ...eep sounds 4 Plug the motor handpiece into the charger with it facing forward The charging lamp lights up and charging starts 5 Orange light is on during charging and green light is on when fully char...

Page 11: ...act dealer The temperature of motor handpiece is high The temperature of battery is high The charger does not work The charge lamp does not light The plug of the charger is not inserted into the outle...

Page 12: ...using a nylon brush after soaking for 20 minutes in Enzymatic Detergent Rinse the angle handpiece using the tap water for 3 minutes to remove Enzymatic Detergent Wash the angle handpiece for 10 minut...

Page 13: ...n lubrication please hold the angle handpiece firmly because the angle handpiece can go out by the spray pressure Do not insert the oil into the motor handpiece It may cause failure When cleaning hand...

Page 14: ...2 4V Approx 2 5 hours Approx 9 hours No load PRODUCT SPECIFICATION Model name Input Output Power consumption TRAUS XEW10 AC100 240V 50 60HZ 0 1A DC 4 5V 7 5VA Model name Gear ratio Bur replacement ty...

Page 15: ...he other device s are connected Consult the manufacturer or filed service technician for help Electromagnetic Emissions The TRAUS ENDO is intended for use in the electromagnetic environment specified...

Page 16: ...t 8 angles 0 UT for 1 cycle 70 UT for 25 30 cycles 0 UT for 250 300 cycle 8kV contact 15kV air 2kV for power supply lines 1kV for input output lines 1kV differential mode 2kV common mode 0 UT for 0 5...

Page 17: ...FM radio broadcasts and TV broadcasts cannot be determined exactly by theory To assess the electromagnetic environment due to fixed radiofrequency emitters an electromagnetic environment measurement...

Page 18: ...1 At 80 MHz and 800 MHz the higher frequency range applies Note 2 These specifications may not be applicable in all situations Electromagnetic propagation is affected by absorption and reflection fro...

Page 19: ...voltage Drop the device during setting moving using Use the not recommended consumables or accessories Malfunction after repairing the product at the not recommended repair shop Normal wear of the con...

Page 20: ...580 0918 Tel 82 53 587 2341 Fax 82 53 580 0999 www saeshin com traus saeshin com This product is medical device MERIDIUS MEDICAL EUROPE LIMITED Unit 3D North Point House North point Business Park New...

Reviews: