Recommended separation distances between portable and mobile RF

communications equipment and the TRAUS ENDO

The TRAUS ENDO is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled.

The customer or the user of the TRAUS ENDO can help prevent electromagnetic interference by maintaining a minimum

distance between portable and mobile RF communications equipment (transmitters) and the TRAUS ENDO as recommended

below, according to the maximum output power of the communications equipment

Rated maximum output

Power of transmitter

(W)

For transmitters rated at a maximum output power not listed above, the recommended separation distance ‘d’ in meters (m) can

be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of

the transmitter in watts (W) according to the transmitter manufacturer.

Note 1

: At 80 MHz and 800 MHz, the higher frequency range applies.

Note 2

: These specifications may not be applicable in all situations. Electromagnetic propagation is affected by absorption and reflection

from structures, objects and persons.

150 kHz ~ 80 MHz

d = 1.2 √P

Separation distance according to frequency of transmitter (m)

800 MHz ~ 2.7 GHz

d = 2.3 √P

0.01

0.12

0.12

0.23

0.1

0.38

0.38

0.73

1

1.2

1.2

2.3

10

3.8

3.8

7.3

100

12

12

23

80 MHz ~ 800 MHz

d = 1.2 √P

Cables and accessories

Maximum length

Shield

Complies with

AC power cord

1.8 M

Unshielded

RF emissions, CISPR11

Class A / Group 1

Harmonic emissions

IEC 61000-3-2

EN 61000-3-2

Voltage fluctuations / flicker emissions

IEC 61000-3-3

EN 61000-3-3

Electrostatic discharge (ESD)

IEC 61000-4-2

EN 61000-4-2

Electrical fast transient / burst

IEC 61000-4-4

EN 61000-4-4

Surge

IEC 61000-4-5

EN 61000-4-5

Voltage dips, short interruptions

and voltage variations on

power supply input lines

IEC 61000-4-11

EN 61000-4-11

Power frequency (50/60Hz)

magnetic field

IEC 61000-4-8

EN 61000-4-8

Conducted RF

IEC 61000-4-6

EN 61000-4-6

Radiated RF

IEC 61000-4-3

EN 61000-4-3

16

Summary of Contents for Traus ENDO

Page 1: ...www saeshin com INSTRUCTION MANUAL ENDO...

Page 2: ...N PRODUCT SPECIFICATION ELECTROMAGNETIC COMPATIBILITY INFORMATION WARRANTY SYMBOL Table of contents This product is cordless handpiece for root canal treatment This motor has to be used only in hospit...

Page 3: ...pment Type of protection against electric shock Class II equipment Degree of protection against electric shock Type BF applied part Classification according to the degree of protection against ingress...

Page 4: ...tion the handpiece correctly Pushing it into the charger forcibly in the incorrect direction may cause damage Take care not to put needle pin into the charging terminal of the battery charger Place th...

Page 5: ...lace If the batteries leak do not allow the liquid to come into contact with your skin eyes or clothes If the batteries leak from the motor handpiece or there change of the exterior or part discolor o...

Page 6: ...1 10 1 16 1 20 1 32 1 The free running speed of the angle handpiece shall be in accordance with the manufacturer s instructions at a tolerance of 10 as specified Refer to ISO 14457 Prophy products mus...

Page 7: ...lied to the motor it may automatically step to prevent overheating In this case leave the motor for a while until it cools down Please check all components are included PRODUCT COMPONENTS PART NAME OF...

Page 8: ...display on the screen or 2 Battery display Shows 3 steps for charging amount 3 Beep display Beep sound function each operating condition 4 Program display Shows the program number currently selected...

Page 9: ...ile inwards by matching the home position until the angle handpiece clicks Disconnection Press the button on the head to pull out the file Turn off the power when insert or remove the file Pull out th...

Page 10: ...eep sounds 4 Plug the motor handpiece into the charger with it facing forward The charging lamp lights up and charging starts 5 Orange light is on during charging and green light is on when fully char...

Page 11: ...act dealer The temperature of motor handpiece is high The temperature of battery is high The charger does not work The charge lamp does not light The plug of the charger is not inserted into the outle...

Page 12: ...using a nylon brush after soaking for 20 minutes in Enzymatic Detergent Rinse the angle handpiece using the tap water for 3 minutes to remove Enzymatic Detergent Wash the angle handpiece for 10 minut...

Page 13: ...n lubrication please hold the angle handpiece firmly because the angle handpiece can go out by the spray pressure Do not insert the oil into the motor handpiece It may cause failure When cleaning hand...

Page 14: ...2 4V Approx 2 5 hours Approx 9 hours No load PRODUCT SPECIFICATION Model name Input Output Power consumption TRAUS XEW10 AC100 240V 50 60HZ 0 1A DC 4 5V 7 5VA Model name Gear ratio Bur replacement ty...

Page 15: ...he other device s are connected Consult the manufacturer or filed service technician for help Electromagnetic Emissions The TRAUS ENDO is intended for use in the electromagnetic environment specified...

Page 16: ...t 8 angles 0 UT for 1 cycle 70 UT for 25 30 cycles 0 UT for 250 300 cycle 8kV contact 15kV air 2kV for power supply lines 1kV for input output lines 1kV differential mode 2kV common mode 0 UT for 0 5...

Page 17: ...FM radio broadcasts and TV broadcasts cannot be determined exactly by theory To assess the electromagnetic environment due to fixed radiofrequency emitters an electromagnetic environment measurement...

Page 18: ...1 At 80 MHz and 800 MHz the higher frequency range applies Note 2 These specifications may not be applicable in all situations Electromagnetic propagation is affected by absorption and reflection fro...

Page 19: ...voltage Drop the device during setting moving using Use the not recommended consumables or accessories Malfunction after repairing the product at the not recommended repair shop Normal wear of the con...

Page 20: ...580 0918 Tel 82 53 587 2341 Fax 82 53 580 0999 www saeshin com traus saeshin com This product is medical device MERIDIUS MEDICAL EUROPE LIMITED Unit 3D North Point House North point Business Park New...

Reviews: