■ Error and remedy
■ Problem and remedy
* When trouble is found, check the following before contacting the manufacturer. If none of these are
applicable or the trouble is not remedied even after action has been taken, please contact the manufacturer.
Error code
Cause & Remedy
Cause: Torque is overloaded than the setting values during the procedure.
Remedy: Press the Motor ON/OFF button to release the error.
Contact the manufacturer if the remedies do not activated.
Problem
Remedy
Cause
Charge the battery.
The power of
motor handpiece
is not turned on
Battery has been discharged.
It is normal that the batteries become a little bit
warm right after charging.
If the batteries are hot while normal operating
conditions, there may be an abnormality.
Contact dealer.
The temperature of
motor handpiece
is high.
The temperature of
battery is high.
The charger
does not work.
(The charge lamp
does not light.)
The plug of the charger is
not inserted into the outlet, or
there is no electricity in the outlet.
Internal fuse has burnt.
Check the connections.
Contact dealer.
The motor handpiece
does not rotate.
Head of contra angle was stuck.
Bonding of motor gear
was detached.
Clean the head or change it.
Contact dealer.
Insert the battery.
Battery has not been inserted.
- The full charge time is about 2.5 hours, but it may be different depending on the battery usage, new and old batteries,
and general operating environment (temperature and humidity). Older batteries may have a shorter charging life.
- While charging, the battery part may get slightly heated, but this is not a malfunction.
- Charging completion time may vary depending on the remaining battery charge.
- Charge the handpiece during the motor resting time.
9