background image

27

If “Alarm at:” is checked, the analog display will turn red and an audible indicator will be played if the count 

rate exceeds the value contained in the input box. The alarm will reset when the count rate falls below the 

alarm level, or when the box is unchecked.
“Meter Response Time” changes the number of seconds of data over which the analog display is averaged. 

Higher values will result in a “smoother” meter action, but slower response to changes in the count rate. This 

can be varied between 1 and 60 seconds.
If “Autorange” is checked, the meter scale will change automatically based on the count rate.
The “Scale Max” changes the base value for the upper limit of the analog display.
The “x 1” through “x 10,000” are scale ranges for the analog display.
“Ratemeter units” allows you to change the analog display between units of counts per minute (cpm) or counts 

per

Procedures

Calibrating a Detector with URSA-II

In order to use URSA-II with a given detector, the following must be done (in this order):
1. Adjust Hardware Settings.
2. Perform Energy Calibration.
3. Perform a Shape Calibration. Following this step, URSA-II is ready to provide qualitative analysis.
4. Create a “Regions of Interest” (ROI) Set. This is in preparation for efficiency calibration, which must be 

coupled to an ROI set.
5. Perform Efficiency Calibration. Once an efficiency calibration has been performed, URSA-II is ready to 

perform quantitative analysis.

Adjust Hardware Settings

Hardware settings are used to adjust input and polarity, high voltage (HV), threshold, gain, fine gain, and 

shaping time.
Hardware settings are unique to each URSA-II /detector combination. The specific settings are loaded when 

a specific detector is selected. If you intend to use the same detector at different hardware settings, you can 

define separate detectors with different descriptions.
Since URSA-II was designed to work with essentially any radiation detector (or more specifically, in the case 

of URSA-II MCA, essentially any gamma detector), the software cannot make any assumptions regarding the 

operating characteristics of the detector and therefore cannot contain automated hardware adjustment. 

Discovering the appropriate hardware settings for a particular detector is essentially a trial-and-error process. 

Summary of Contents for URSA II

Page 1: ...1 S E International Inc P O Box 39 436 Farm Rd Summertown TN 38483 USA 1 800 293 5759 1 931 964 3561 Fax 1 931 964 3564 www seintl com radiationinfo seintl com INTERNATIONAL OPERATION MANUAL...

Page 2: ...s 9 Misc 9 View Menu 9 Sound 10 Library 10 Transfer 10 Help 10 Button Bar 11 Hardware Status Bar 12 New User Hints Bar 12 Spectrum Display 12 Spectrum Status Bar 12 Acquisition Controls and Indicators...

Page 3: ...erforming an Energy Calibration 26 Performing a Shape Calibration 28 Creating a Regions of Interest ROI Set 29 Manually Creating or Editing an ROI Set Manual Method 30 Performing an Efficiency Calibra...

Page 4: ...URSA II accepts positive negative or pre shaped positive pulses Pulse heights should not exceed approximately 5 volts Uses 110V or 220VAC to 12V DC adapter 1 25A or runs on 6 internally recharged NiM...

Page 5: ...ined batteries voltages The NiMH batteries should power the URSA II for about four hours Note however that this figure can change dramatically based on the current draw of the detector high voltage ot...

Page 6: ...in and power on URSA II 4 Start the software URSA II MCA searches all COM ports for the URSA II automatically 5 A detector must be entered before URSA II can be opened Many of URSA II s calibration f...

Page 7: ...ing a displayed spectrum to an URSA II spectrum file usf The next two Save Spectrum and Load as Background and Save Spectrum and Load to Static Display also save the spectrum but additionally either l...

Page 8: ...order to save printer ink This is the recommended method of printing charts The final menu item Exit closes URSA II MCA Warnings will occur if the currently displayed spectrum has not been saved and...

Page 9: ...n See instructions on page 32 The next seven QQQ menu items Full Screen through Sample Data Panel are identical in function to the seven buttons in the Button Panel on the lower right of the Spectrum...

Page 10: ...h nn ss where yyyy is the four digit year mm is the two digit month dd is the two digit day of the month hh is the two digit hour in 24 hour format nn is the two digit minutes past the hour and ss is...

Page 11: ...se common sense in applying different colors For example if the Background color is set to be the same as the Main Spectrum color the display will appear blank Undo Zoom gives a method for restoring t...

Page 12: ...button caption becomes Hide and clicking on it or on the x in the upper right corner will close the window The order of the display can be based on energy or yield ascending or descending Change this...

Page 13: ...area Once a spectrum is zoomed it can be panned by rightclicking the mouse on the spectrum display and dragging it Un zoom by dragging a box any size anywhere on the spectrum display from lower right...

Page 14: ...isition has previously been stopped The Clear button erases all data from the spectrum and resets the elapsed time to zero If a spectrum is currently acquiring acquisition will stop No warning is give...

Page 15: ...when the spectrum is displayed in 256 or so channels URSA II MCA s 4096 channels of data can be filtered to give a smoother representation of the energy continuum Try using different values of smoothi...

Page 16: ...nudges to compensate for spectrum drift caused by temperature changes in the detector If an even smaller shift in gain is needed try adjusting the high voltage by tiny increments including decimal val...

Page 17: ...set This name becomes the one displayed in the spectrum status bar and is also used to identify individual ROIs in analysis reports Approximately in the middle of the ROI panel are two radio button co...

Page 18: ...will be unlocked for editing If this box is not checked the data for the last calibration can be viewed but not altered The arrow buttons will change the currently selected specific ROI within the gre...

Page 19: ...ed printed or saved by clicking the Report button Note The contents of the report and use of the report window are discussed in detail in other portions of this manual Peak Based This displays the Pea...

Page 20: ...h and this method of adding found peaks may beused instead of or as a supplement to the computerized search When Show FWHM is checked the FWHM full width at half maximum calculated for each found peak...

Page 21: ...sults will be displayed in subsequently displayed reports The Save Spectrum button repeats the function under the File Menu It is also included here for convenience The Clear Fields button deletes the...

Page 22: ...emoved to improve performance Both libraries contain many calculated multiplets as an aid These can of course be edited or disabled Either library can be edited and saved under a different name creati...

Page 23: ...ently displayed isotope Click on any photon energy in the grid and use the Kill Energy button to completely delete that photon energy from the currently displayed isotope Use the Add Isotope button to...

Page 24: ...g or recording changes in count rate over time An analog meter style display also shows the current count rate MCS mode allows you complete control over how the data is accumulated and displayed Menu...

Page 25: ...rt trace to be displayed as a line as points as a line with points the default bars or as an area graph The next three menu items Chart Background Color Chart Trace Color and Chart Labels Color allow...

Page 26: ...unt of time in the input box Typing in the Time box will immediately un check it pressing Enter on the keyboard after changing the Time value will re check it Directly underneath the Time sec input bo...

Page 27: ...erform Energy Calibration 3 Perform a Shape Calibration Following this step URSA II is ready to provide qualitative analysis 4 Create a Regions of Interest ROI Set This is in preparation for efficienc...

Page 28: ...e URSA II has two detector inputs Input 1 is the series C connector right next to the DB9 serial connector This input has the signal and the high voltage coupled just like on most hand held instrument...

Page 29: ...it can be displayed usefully as a spectrum These two controls are used together to create an overall gain The overall gain is generally the value you should follow It is worth mentioning that the gai...

Page 30: ...ed If different sets of hardware settings are desired for the same detector create a new detector based on that detector changing the description to reflect the different purpose Procedure 4 Click on...

Page 31: ...that detector is attached to the URSA II the URSA II is switched on the URSA II MCA software has been successfully started a radioactive source of known isotopic composition is on the detector and URS...

Page 32: ...simple spectrum such as 137Cs 2 From the Utility menu select Energy Calibrate The energy calibration window will appear If this is the first energy calibration for this detector the energies for the...

Page 33: ...een performed Discussion For NaI Tl detectors the Full Width at Half Maximum FWHM is proportional to the square root of the centroid energy as is the overall peak area The Shape Calibration helps dete...

Page 34: ...t method a spectrum is acquired and the Find Peaks buttons is pressed from the Peak Based panel edited as desired then saved as an ROI set The ROI set can be further edited from the ROI Panel includin...

Page 35: ...while the single arrows adjust the settings in approximately 0 5 keV increments c To delete a peak first select the peak to be deleted using the Selected Peak arrow set Once the peak is highlighted cl...

Page 36: ...io button c Position the cursor at the suitable location on the spectrum display and double click d The energy and associated channel will be displayed in the appropriate boxes on the ROI Panel e When...

Page 37: ...sed Procedure 1 Acquire a spectrum of your efficiency calibration source A background spectrum should also be subtracted from the display if that is how your samples will be analyzed 2 Click on Effici...

Page 38: ...small window will pop up and display an isotope selection drop down menu and an energy peak selection drop down menu The hardware settings panel will also open along the bottom of the spectrum display...

Page 39: ...ts suppliers be liable for any consequential incidental direct indirect special punitive or other damages whatsoever including without limitation damages for loss of business profits business interrup...

Page 40: ...or will repair the defect and return the instrument to you prepaid without charge for parts or labor Shipping cost to the factory for repair is the responsibility of the customer Customer is also resp...

Page 41: ...S CITY STATE ZIP COUNTRY Please fill out this form and send it back to us if you would like to be notified of the NIST calibration renewal for your instrument to S E International Inc P O Box 39 436 F...

Reviews: