7.4 

Data Transfer via Modbus Interface

The con::cube is equipped with a Modbus interface which support the standardized Modbus protocol via RTU or TCP / IP. This 

interface enables the integration of the con::cube into other networks and supports the following features:

  Reading of device description (see table - Mapping of Device specific Input Register or Device specific Holding Register)

  Reading of parameter results (see table - Mapping of Parameter specific Input Register or Parameter specific Holding 

Register)

  Reading of parameter status information (see table - Mapping of Parameter Status)

  Reading of system status information (see table - Mapping of System Status)

Access to this information is possible by reading input registers and / or by reading holding registers. The type and the location 

of the registers is described in the following sections.

The con::cube Modbus is specified according the Modbus-IDA Application Protocol V1.1a and supports the following function 

codes:

Function

Function code

Description (used for)

Read Input Register

0x04

Parameter readings

Parameter status

Device status

Read Holding Register

0x03

Device and parameter configuration settings, 

Mirrored input registers (for support of limited Modbus 

master devices) 

M a p p i n g   o f   D e v i c e   s p e c i f i c   I n p u t   R e g i s t e r

Input register,

16bit, RO

Tag name

Address

Type

Length R/W

Description

Address

device

description

uiVersion

0x0000

unit16

1

R

Version of Modbus mapping 

protocol. For all changes in public 

registers: 0xAABB

AA ... Major version

BB ... Minor version (compatible) 

0

eVendor

0x0001

enum

1

R

Vendor code

1

eModel

0x0002

enum

1

R

Device model

2

abModel

0x0003

char[20]

10

R

Description of device model, 

filled with spaces

3

abSerialNumber 0x000D

char[8]

4

R

Serial number, 

filled with spaces

13

abSWRelease

0x0013

char[4]

2

R

Software release: 0xAABB

AA ... Major version

BB ... Minor version

19

device status 

public

tSampleTime

0x0068

timestamp 6

R

Time when the parameter results 

have been updated. Timestamp of 

logged status and results

104

bmDeviceStatus 0x0078

bitmask

1

R

Device status

120

con::cube D-330, 01-2020 Release

www.s-can.at

Copyright © s::can Messtechnik GmbH

25 / 42

Summary of Contents for con::cube D-330

Page 1: ...Manual con cube D 330 January 2020 Release S 45 M...

Page 2: ......

Page 3: ...Mode 16 6 Operation of con cube 17 6 1 Direct Operation of con cube 17 6 2 Remote Operation of con cube 17 6 2 1 Remote Operation of con cube via Web Browser 17 6 2 2 Remote Operation of con cube via...

Page 4: ...Cleaning 35 11 2 1 Cleaning Valve 35 11 3 Spare Parts 35 11 3 1 Desiccant Package 35 11 3 2 Touch Pen 35 11 4 Optional Features 36 11 4 1 Gateway to 3G Modem 36 11 4 2 Analog Output Module 36 11 4 3 D...

Page 5: ...d in English and German by s can This original manual serves as the reference in case discrepancies occur in versions of the manual after translation into third languages This manual and all informati...

Page 6: ...products are leaving our factory in immaculate technical and safety conditions Inappropriate or not intended use of the product however can cause danger The manufacturer is not responsible for damage...

Page 7: ...ia analog input optional Integration of external sensor signals via digital input optional Network connectivity via ethernet WLAN or optional 3G 4G modem Remote control of s can monitoring station via...

Page 8: ...on of third party readings input module D 315 IN RELAY 2 digital inputs for integration of third party readings input module D 315 ANTENNA PLUG internal antenna adapter cable and connector D 315 ANTEN...

Page 9: ...le glands M16 optional for Profibus 3 cable glands M16 1 cable gland M16 optional antenna plug 2 cable glands M20 Holder for housing cover Desiccant package Insert for SIM card Additional cable termin...

Page 10: ...s well as electromagnetic radiation Damage to the device caused by wrong storage will not be covered by warranty Transport should be done in a packaging that protects the device original packaging or...

Page 11: ...an option Cable for power supply part no C 31 xx Gateway to 3G part no D 315 3GLX Internal antenna adapter cable and connector part no D 315 ANTENNA PLUG External antenna part no D 315 ANTENNA PRO Di...

Page 12: ...Best possible weather and splash water proof set up Infrastructure energy data and compressed air Power supply for controller for operation operational reliability voltage power peak free Oil and part...

Page 13: ...1 Plan of Terminal Connections Once the housing cover has been opened four hexagonal head screws removed you have access to the cable terminal compartment of the con cube Opening of the cable terminal...

Page 14: ...ox B 23 xxx has to be used to connect a further spectrometer probe 5 3 Connection of s can ISE Probes and Sensors Alternative or additional to the spectrometer probe it is possible to connect ISE prob...

Page 15: ...arameterised within the moni tool software please refer to manual moni tool 5 5 Connection of Main Power Supply This type of work must be performed by authorised persons only see section 2 Depending o...

Page 16: ...display Please refer to the manual of the operation software moni tool for all further tasks to complete the initialisation of your s can monitoring system 5 7 con cube Sleep Mode For operation with l...

Page 17: ...6 1 Direct Operation of con cube Normally this direct operation is performed by a person standing in front of the con cube and using the touch screen Alternatively moni tool can also be operated with...

Page 18: ...once the remote connection was successful The password to be entered is scan case sensitive If you operate the con cube via VNC Viewer all actions can be seen on the touch screen of the con cube also...

Page 19: ...contains all types of possible connectivity setups Regarding the individual connection types please refer to the subsections afterwards con cube Type of data transfer Connection type Customer infrast...

Page 20: ...cube 192 168 10 12 IP address of the PC 192 168 10 2 Subnet mask for both devices 255 255 255 0 Connect your network cable RJ45 to the con cube and to the PC when working via LAN see section 6 5 1 Al...

Page 21: ...ofile e g by editing the existing Default profile via Service Terminal Network Settings Modem Select Default and push icon Edit Enter a Profile name and the PIN of the SIM card Enter the Apn and the A...

Page 22: ...ich is always OpenVPN Custom Push the button Save Within the next window enter a Profile name Select the file containing the Private key password e g password txt Select the file containing the OpenVP...

Page 23: ...moni tool menu item Service Outputs TML Transfer of measurement result files automatical via FTP see manual moni tool Service Outputs File Output FTP Transfer automatical via SSH see manual moni tool...

Page 24: ...eger 0xXXYY int16 has the following mapping M o d b u s r e g i s t e r Register address b15 b14 b13 b12 b11 b10 b9 b8 b7 b6 b5 b4 b3 b2 b1 b0 0xXXXX 0xXX 0xYY For example 17289 0xBC77 has the mapping...

Page 25: ...ion Function code Description used for Read Input Register 0x04 Parameter readings Parameter status Device status Read Holding Register 0x03 Device and parameter configuration settings Mirrored input...

Page 26: ...s 0x00C8 bitmask 1 R Para 10 status 1 200 xP10Value 0x00CA float 2 R Para 10 result 202 parameter 11 result bmP11Status 0x00D0 bitmask 1 R Para 11 status 1 208 xP11Value 0x00D2 float 2 R Para 11 resul...

Page 27: ...0x0162 float 2 R Para 29result 354 parameter 30 result bmP30Status 0x0168 bitmask 1 R Para 30 status 1 360 xP30Value 0x016A float 2 R Para 30 result 362 parameter 31 result bmP31Status 0x0170 bitmask...

Page 28: ...not supported or reserved Each bit reperesents a condition 1 error 0 o k M a p p i n g o f P a r a m e t e r S t a t u s b m P x S t a t u s Bit Display xPxValue Status Description b0 NaN or value Ge...

Page 29: ...atus Status of parameter 7 0x2A 0x2D Float xP7Value Reading of parameter 7 0x2E 0x2F Bitmask bmP8Status Status of parameter 8 0x30 0x33 Float xP8Value Reading of parameter 8 0x34 0x40 TAI64n tSampleTi...

Page 30: ...remind that polling of readings needs several seconds Automatic measurement active Readings are actualized Service mode not active Check measuring settings Leave Service mode System status LED is blu...

Page 31: ...have to be checked for cleanliness possible damage and dirt or foreign bodies before closing the housing cover every time In case of damage to the cord gasket 2 in the housing cover it has to be repai...

Page 32: ...rn Material Authorization Return consignments of the s can monitoring system or parts of the system shall be done in a packaging that protects the device original packaging or protective covering if p...

Page 33: ...2 x M8 U bolt with screw nut and spring washer 4 x M5 flange nut included in delivery for mounting onto rails 11 1 2 Power Supply Cord For connection of the device to the power supply a power supply...

Page 34: ...install antenna under large overhanging tree branches The mounting of the antenna is performed by the following steps Fix carrier brackets on mounting console with bolts Fix mounting console antenna...

Page 35: ...2 coupler screw connection 6 4 mm nickel plated brass P inlet pipe compressor compressed air side Connection fitting probe side B 44 Hose ID 3 8 inch B 44 2 plug in fitting 6 mm OD A inlet pipe probe...

Page 36: ...see sec tion 5 1 for correct wiring Configuration via moni tool software see section 12 1 for address mapping Assembling only by qualified personal authorized by s can con cube can be equipped with ma...

Page 37: ...pped wire via terminal block see sec tion 5 1 for correct wiring Configuration via moni tool software see section 12 1 for address mapping Assembling only by qualified personal authorized by s can con...

Page 38: ...Hz wide range power supply unit Power supply cable outside diameter 5 12 mm single cable wires 1 5 2 5 mm2 min 80 C min 300 VAC Required galvanic isolation of power input 3 0 kV 1 5 kV D 330 230 D 33...

Page 39: ...ax 8 parameters Analog OUT 4 to 20 mA active Resolution 12bit 5 A Accuracy 30 A Max load 500 Ohm Galvanic isolation 1 kV D 315 OUT PROFIBUS blue module address 1 see section 11 5 4 D 315 OUT SDI12 see...

Page 40: ...ot to be used Slot 2 51 and 52 19 and 20 35 and 36 Slot 3 53 and 54 21 and 22 37 and 38 Slot 4 55 and 56 23 and 24 39 and 40 Slot 5 57 and 58 25 and 26 41 and 42 Slot 6 59 and 60 27 and 28 43 and 44 S...

Page 41: ......

Page 42: ...scan Messtechnik GmbH Brigittagasse 22 24 1200 Vienna Austria Tel 43 0 1 219 73 93 0 Fax 43 0 1 219 73 93 12 office s can at www s can at...

Reviews: