![s::can con::cube D-330 Manual Download Page 29](http://html1.mh-extra.com/html/s-can/con-cube-d-330/con-cube-d-330_manual_3128382029.webp)
7.5
Data Transfer via Profibus DP Interface
The Profibus DP output module (D-315-OUT-PROFIBUS) provides a Profibus DP compatible fieldbus interface (according
to IEC 61158) for the integration of the con::cube into Profibus DP networks. Profibus DP supports up to 244 Bytes of output
data. This enables the transfer of readings and status information for up to eight parameters as well as the system status. The
location and type of the values are shown in the table below.
Profibus Position
Data Type
Name
Description
0x02 – 0x03
Bitmask
bmDeviceStatus
System status of con::cube
0x04 – 0x05
Bitmask
bmP1Status
Status of parameter 1
0x06 – 0x09
Float
xP1Value
Reading of parameter 1
0x0A – 0x0B
Bitmask
bmP2Status
Status of parameter 2
0x0C – 0x0F
Float
xP2Value
Reading of parameter 2
0x10 – 0x11
Bitmask
bmP3Status
Status of parameter 3
0x12 – 0x15
Float
xP3Value
Reading of parameter 3
0x16 – 0x17
Bitmask
bmP4Status
Status of parameter 4
0x18 – 0x1B
Float
xP4Value
Reading of parameter 4
0x1C – 0x1D
Bitmask
bmP5Status
Status of parameter 5
0x1E – 0x21
Float
xP5Value
Reading of parameter 5
0x22 – 0x23
Bitmask
bmP6Status
Status of parameter 6
0x24 – 0x27
Float
xP6Value
Reading of parameter 6
0x28 – 0x29
Bitmask
bmP7Status
Status of parameter 7
0x2A – 0x2D
Float
xP7Value
Reading of parameter 7
0x2E – 0x2F
Bitmask
bmP8Status
Status of parameter 8
0x30 – 0x33
Float
xP8Value
Reading of parameter 8
0x34 – 0x40
TAI64n
tSampleTime
Time of last measurement
7.6
Data Transfer via SDI12 Interface
The SDI12 module (D-315-OUT-SDI12) provides a SDI12 compatible fieldbus interface for the integration of the con::cube into
SDI12 networks. The module enables the transfer of readings and status information for up to eight parameters as well as the
system status. The location and type of the values are shown in the table in section 7.5.
Parameter readings can be queried via a SDI12-Master connected to the con::cube using a simple terminal programm (e.g.
Docklight). For reading of measurement results the following commands are needed (ASCII-code):
Action
Command
Description
Send
0M!<CR><LF>
Start measurement parameter 1
Answer
00001<CR><LF>
Send
0D!<CR><LF>
Query of measurement result
Answer
Measurement result of parameter 1
Send
0M1!<CR><LF>
Start measurement parameter 2
Answer
00001<CR><LF>
Send
0D!<CR><LF>
Query of measurement result
Answer
Measurement result of parameter 2
...
Send
0M7!<CR><LF>
Start measurement parameter 8
Answer
00001<CR><LF>
Send
0D!<CR><LF>
Query of measurement result
Answer
Measurement result of parameter 8
<CR> ..... Carriage Return
<LF> ..... Line feed
con::cube D-330, 01-2020 Release
www.s-can.at
Copyright © s::can Messtechnik GmbH