15
Cobia Flex R/F - User Manual 2014.3A
mAs measuring socket
Cobia MAS Cable
High Voltage X-ray Transformer
Tube current ground return line
4. Connect the other end of the Cobia MAS Cable to the MAS input of the
Cobia Flex R/F.
Caution!
Do not forget to replace the mAs socket jumper when
removing the Cobia MAS Cable.
3.4.2
Tube Current Safety Precautions
Only use the Cobia Internal MAS if you are authorized and skilled to make
invasive measurements using the X-ray generator mAs measuring socket.
Please be aware that high voltage may occur on the Cobia Internal MAS if
the cable connection to the X-ray generator mAs socket is broken. Always
wait approximately one minute after switching off the X-ray generator
before connecting or disconnecting the cables between the Cobia Internal
MAS and the mAs measuring socket.
Always check that the short circuit path (jumper) normally connected to
the mAs measuring socket is reapplied after the use of the Cobia Internal
MAS and before the X-ray generator is switched on again.
Important!
Do NOT connect the Cobia Internal MAS to the filament current circuitry.
The Cobia Internal MAS is only intended to measure the tube mA and