Cobia Flex R/F - User Manual 2014.3A
14
4. Start an exposure.
5. When the measurement is over, the recorded values are by default
cycled on the Cobia display. Press the Right arrow button to go to the List
view. All measurement results from the internal MAS input is marked
with a MAS symbol (shown below).
6. Record the results.
7. To repeat, change settings and start over from Step 5.
3.4.1
Connecting the MAS Cable
Please be aware that high voltage may occur if the mAs probe is not
properly connected. Read the
Tube Current Safety Precautions
section
before starting the measurement.
1. Power down the X-ray generator and wait the recommended time to
allow for generator discharge according to generator manufacturer's
specifications.
2. Remove the jumper of the mAs measuring socket of the generator.
3. Connect the Cobia MAS Cable to the mAs measuring socket using the
included safety cables, observing the polarity (see picture below). If
needed, use included adapters.
15