11.11.5 - Small air bubbles rising in pipette
Usually this is caused by a dirty or damaged fill nozzle.
▪
Observe the maintenance schedules.
▪
Clean the fill nozzle.
▪
Check the fill nozzle for damage. If necessary, replace the fill nozzle.
Sample tube is leaking on the fill nozzle side.
▪
Replace the silicon sample tube
Small air bubbles result in ESR Error 3.
11.11.6 - Random air bubbles in pipette
1. Check Diluent flow by priming the diluter system.
2. Insufficient sample volume.
Random air bubbles result in ESR Error 3.
11.12 - Quality control troubleshooting
Error messages
E115: QC expired, not sampled!
The used Starrsed Control is out of date, no ESR result is
given
▪
Check expire date
▪
Use a new batch of Starrsed Control
E116: QC is out of acceptable range!
Troubleshooting