Run Fill and clean
X
X
X
X
Replace Waste cassette assembly
X
X
X
ESRI090921
4
Replace Blotting washer waste pump
X
X
X
ESRI090026
4
Replace Pinch valve tube
X
X
X
ESRI010246
4
Replace Blue disc filter
X
X
X
QWLV040003
4
Check needle condition
X
X
Replace teflon tips (diluter syringe repair
set)*
X
X
QWLV030901
2
Replace Waste pump motor
X
ESRI090920
1
Replace Fill block washer
X
ESRI030906
1
Replace Sample probe assembly
X
ESRI050909
1
Replace Pipette valves bodies (84 pieces)
X
QTST040001
84
Replace Outer needle assembly
X
ESRI050950
1
Replace all tubing (with Tubing set)
X
EHST079200
1
Compression spring outer needle
X
EHST050053
1
Needle cable assembly
X
EHST089056
1
*Only applicable for Diluter syringe QWLV030009, Diluter syringe QWLV030021 has to be replaced after 2 years
Note:
Periodical Maintenance Kit SEHST110985, containing all parts for yearly maintenance
Two years Periodical Maintenance Kit: SESRI120906
Appendix for Interrliner
MRN-164_2-EN Version 2.02 Interrliner
105