1. Go to tab
[Maintenance]
and perform the
[End-of-day wash]
procedure.
2. Check system for leakage.
▪ Inspect the peristaltic pump tubes and connections for leaks.
▪ Check that liquid does not run back after the pumps have stopped.
3. Clean the outer needle with disinfectant
4. Check tubing from the syringe for trapped air bubbles.
5. Check Diluent syringe for trapped air bubbles.
6. If trapped air bubbles are found, go to tab
[Maintenance],
click button
[Prime / Clean]
and perform the
[Prime
Diluent/Diluter]
function.
7. Wipe outer surface and stainless steel plate below the pipettes with disinfectant.
8. Check (or replace if needed) two drip trays (ESRI010318)
Exchange air filter
Air filter replacement QWLV040003
1. Pull both tube connectors out of the blue disk filter.
2. Place new blue disk filter
3. Reconnect the tube connectors on the filter
Work instruction Interrliner
MRN-164_2-EN Version 2.02 Interrliner