background image

7.3 - Limit error settings

When an option (at Limit error settings) is set to 

YES

 and this limit error occurs, results will be printed/send

to the LIMS. Together with the sedimentation time and dilution rate (which are still printed at the usual

position), the operator/analyst can see what caused the error and may or may not use the ESR values which

are preserved in the error message.
When an option is set to 

NO

 and this limit error occurs, the fields for 

30 min ESR, 60 min ESR and the

temperature corrected ESR are filled with spaces and thus results are not printed/send to the LIMS.
The error message in the error field (column 10) indicates that at least one of the limits (ESR time, dilution

rate, column height, bubbles on top, hazy aspects and temperature) has been exceeded.
Description of the error message

 L_err(hhh/www/ttt/ccc)

 :

L_err

 means it is a "limit error"

hhh

 is the 30 minutes ESR

www

 is the 60 minute ESR

ttt

 is the temperature corrected 60 minute result

ccc

 is the column height

Example of a limit error message:

L_err( 42/ 84/ 75/200) means 42 mm in the 30 minute method and temperature correction 75 with a

correct column height.

L_err(---/ 84/ 75/200) means 84 mm in the 60 minute method and temperature correction 75 with a

correct column height.

Note:

 If limit error filters in the LIMS are not corresponding with the limit error settings on the instrument it

is possible that not all expected results are send to the LIMS.

7.4 - Reporting range

The reporting range in the columns 2, 3 and 4 are in millimeters. The start of the measure range is at the

top of the meniscus down to 140 mm. If the detection of cells/plasma is over 140 mm then the report will be

>140.

7.5 - Aspect Hazy

The automatic reading of the Westergren sedimentation pipettes is performed by moving an optical sensor

along the pipettes. While the sensor is moving, it is taking a reading every 0.25 mm. The sensor is reading

the absorption of infrared light through the Westergren pipette filled with blood. From these readings, values

at a number of absorption levels are determined. All absorption figures are relative to the darkest and

lightest reading (darkest = 100 % and the lightest = 0 % absorption respectively).
By definition the levels are:

87.5%

Cells/ plasma separation

75.0%

Hazy detection

50.0%

Meniscus detection

Reporting

MRN-164_2-EN Version 2.02 Interrliner

65

Summary of Contents for EHST109621

Page 1: ...Instructions for Use Interrliner MRN 164_2 EN Version 2 02...

Page 2: ......

Page 3: ...ciple 13 2 Instrument description 14 2 1 PC Operation and User Interface 15 2 2 Technical specifications 16 2 3 Used reagents 19 2 3 1 Starrsed Rinse solution 19 2 3 2 Starrsed Saline 19 2 3 3 Starrse...

Page 4: ...ple history QC 41 5 3 5 6 Linked QC IDs 41 5 3 5 7 QC Result analysis 42 5 3 6 Set start end date 43 5 3 7 History sample analyse option 43 5 3 8 History analyse results 45 5 4 Reagents screen 46 5 4...

Page 5: ...2 Turn off sequence 61 7 Reporting 62 7 1 Result Printout 62 7 1 1 Report 60 Minute mode 63 7 1 2 Report 30 Minute mode 64 7 2 ESR Error 64 7 2 1 ESR Error and Warning code messages 64 7 3 Limit erro...

Page 6: ...Fill time out error 75 11 8 Hazy reports 76 11 9 Leaking pipettes 77 11 10 Liquid level sensor not sensing 77 11 11 Air bubbles 77 11 11 1 Foam in column 78 11 11 2 Pipette looks like zebra crossing 7...

Page 7: ...acement 88 13 3 4 Fill and clean procedure 88 13 4 Level 3 maintenance 90 13 5 Check or replace sample probe or outer needle 91 13 6 Reagents installation 92 13 7 Cleaning Touch screen display 93 14 A...

Page 8: ...al MRN 164_2 is applicable for the following instruments EHST109621 Interrliner XN1 FRL EHST109622 Interrliner XN2 FRL EHST109623 Interrliner XN3 FRL with Software version V6 xx Application 8 MRN 164_...

Page 9: ...andatory H Schavemaker 2 01 jan 2020 Update and publication H Schavemaker 2 00 June 2017 Instrument description Safety Program Maintenance Technical specifications additions Explanation of symbols Int...

Page 10: ...e tubes reduces the possibility of contamination for the user and environment Standard glass Westergren pipettes are used in which the measurement can be corrected to a constant temperature 18 C Celsi...

Page 11: ...Service manual MRN 125 1 4 Explanation of symbols The following symbols can be used in this manual and for the instrument Warning sign to prevent personal injury due to biohazard Warning sign to preve...

Page 12: ...vailable in several languages 2 User Manual UM If available Intended for the lab supervisor Contains information from the IFU and additional information concerning settings service higher maintenance...

Page 13: ...of a sample 1 9 Dilution principle The principle of adding Diluent to a flow of whole blood is unique The Interrliner has the capability of monitoring the air displacement during the aspiration cycle...

Page 14: ...acing successively each sample tube on the ESR analyser unit for aspiration Conveyor belt running from the Start pool Front Return line for transport of racks back to Sysmex HST XN after sampling Prot...

Page 15: ...m The user interface is intuitive and can be activated via the keyboard or the optional touch screen All the data from each sample including the raw measuring data and a pictorial representation of th...

Page 16: ...ical Laboratory World 1994 Allowed blood specimen types For EDTA mode Whole blood with 1 EDTA anticoagulant Automatic dilution 4 vols blood 1 vol Starrsed Diluent Reported result mm after 1 hour Throu...

Page 17: ...0 60 Hz Fuse 20 x 5 mm Per ESR Unit Slow blow 230V 2 5 Amp 115V 5 Amp Transport Unit Slow blow 230V 1 6 Amp 115V 3 15 Amp XN1 XN2 XN3 Power consumption Standby 90 VA 2x90VA 3x90VA Maximum 650 VA 2x100...

Page 18: ...22 kg double 26 triple LIS connectivity Communication Ethernet and serial bi directional according various protocols Data storage Storage medium Hard disk on external PC Storage capacity indication a...

Page 19: ...blood samples in Starrsed ESR analyzers Approximately 0 5 ml Diluent is used for each sample Approximately 2 5 ml is used for one Prime cycle The diluent container is a 5 liter container QRR 010931 2...

Page 20: ...n the cabinet beneath the Start Pool Attention The sample robot unit is powered by the ESR analyzers power supply Switching the transport units OFF will not switch OFF the sample robot A central power...

Page 21: ...ment must be made inoperative and be secured against any unintended operation The matter should then be referred to qualified technicians Safety protection is likely to be impaired if for example the...

Page 22: ...cover is closed and locked To get access to the tube handling mechanism e g for maintenance or other necessary operator intervention Switch the Sample mode OFF by pressing the sample button The sampl...

Page 23: ...creen The Reagents screen is used to check the reagent levels and log reagent replacement To activate priming sequences and cleaning operations the screen Maintenance is used The Settings and Service...

Page 24: ...www rrmechatronics com Login part The following program description is valid for software from version 6 12 1 and up and up 5 2 Sample screen The main menu is displayed during operation To access oth...

Page 25: ...DE the instrument uses the standard saved settings For instance the ESR time is set back to 60 minutes or 30 minutes according the used method The User Manual button is at the bottom of the menu Click...

Page 26: ...r this sample ESR is required and waiting to be done Sample was aspirated successfully Sample was aspirated but with fill errors After processing the rack the information of the rack is transferred to...

Page 27: ...us Information about the current status of the instrument is shown here such as the selected mode EDTA or Citrate selected method 60 or 30 minute and symbols that draw attention to certain maintenance...

Page 28: ...5 2 3 Pipette information The following information is shown Starrsed program 28 MRN 164_2 EN Version 2 02 Interrliner...

Page 29: ...te is marked to be washed when it reaches the rinse station This indicator is also on when a sample could not be aspirated properly Pipette disabled In case of a failing missing or broken pipette indi...

Page 30: ...lected result to the printer Print result header Only if option Settings General settings Print after measurement is switched ON it is possible to print a result header Send all to HOST Send all resul...

Page 31: ...process time After measuring the pipette the pipette data is transferred to the sample history files In the header of the table the names of the columns are shown Double click the header of any column...

Page 32: ...a selected time frame of the day Show results of a number of past days Default value is set for 7 days Show results of a specific day Show results of the range between the first selected date to the...

Page 33: ...ys Show results of a specific day Show results of the range between the first selected date to the next selected date From the selected Sample ID detailed information is shown on this screen Sample ID...

Page 34: ...r Pipette in which the sample was measured Error code Shows any ESR error or warning code e g Too many borders found A complete overview of measurement data is shown with activating the option Display...

Page 35: ...ervice General Barcode the rack number is shown under date time The Clear rack history button will clear the contents of the rack history file and restart to build up a new rack history file 5 3 3 1 D...

Page 36: ...Bubbles on top Increasing samples with bubbles indicate the need for maintenance of the aspiration system see Foam in column ESR errors Increasing ESR errors may indicate the need for maintenance see...

Page 37: ...sted in Patient results Note This part of the software can only be used in combination with Starrsed Control as quality control material 5 3 5 1 QC Normal results table The results from Starrsed Contr...

Page 38: ...sed Control expiry date is exceeded it is not possible to continue with this QC sample The sample is not measured but the failed attempt is logged in the table Expected ESR This is the temperature cor...

Page 39: ...MS is activated at QC Settings Related patient results This screen is simular to the Display sample history screen The background colour of the patient history table is switched to light yellow to dis...

Page 40: ...viation the average deviation of all QC results compared with the expected ESR Coefficient of variation ratio of the standard deviation to the expected ESR expressed in a percentage Number of QC resul...

Page 41: ...g QC result The results are presented in the layout of the Display sample history screen Depending on the frequency of QC samples related patient results may span over multiple days and are listed per...

Page 42: ...Settings the links can be removed automatically when a usable ESR result has been reported for this particular Lab ID 5 3 5 7 QC Result analysis Authorized staff should identify and differentiate acc...

Page 43: ...5 3 6 Set start end date Enter the Start End date and time 5 3 7 History sample analyse option Starrsed program MRN 164_2 EN Version 2 02 Interrliner 43...

Page 44: ...today s results between start time and end time or make a selection for 1 A specific number of past days 2 A specific date 3 A range of days from start date to end date Starrsed program 44 MRN 164_2...

Page 45: ...all the samples with a dilution rate 110 are displayed in the table By selecting Dilution 90 all the samples with a dilution rate 90 are displayed in the table Aspect By selecting one of the three Haz...

Page 46: ...o fill in all reagent information If not filled in correctly the instrument will give an alarm and does not aspirate samples The article number always has to be added this to prevent replacing a reage...

Page 47: ...ory of the used reagents Select the reagent type on the right side For external use of the information all the logged reagent data can by exported to EXCEL CSV format by clicking Export to Excel CSV S...

Page 48: ...me for all reagents Input screen for new reagents Make a selection to add new default setting or delete the current information and continue with Next 5 4 2 1 New reagent input cont Data can be entere...

Page 49: ...he date when the reagent was placed 5 Check if the preview box shows the correct information then press OK 5 5 Maintenance screen When there is a sensor alarm an alarm indicator is shown in the tab Th...

Page 50: ...outer needle sample probe and fill nozzle are washed Prime Diluter The Diluter prime cycle is 5 strokes of the syringe Prime Nozzle with water After each aspiration the fill nozzle is flushed with de...

Page 51: ...onged use of the instrument proteins are building up in the Westergren pipettes which need to be removed using a strong cleaning agent This function fills all pipettes with a cleaning agent and remove...

Page 52: ...er does not start during the aspiration the status of this sensor must be checked Click the Check button the OK sign is shown if the sensor value is in range Check Diluent flow sensor This sensor is o...

Page 53: ...position correct Measure head start position wrong When a test pipette is installed at the measuring position the result of the test pipette is displayed in the field ESR mm When the sensor is out of...

Page 54: ...of the instrument and locate possible problems This log can be saved e g to a memory stick by clicking button Save As Note since software version 6 12 5 and up is a sample counter included in every e...

Page 55: ...an alarm indicator is shown in the tab Maintenance All performed maintenance functions are logged automatically This log can be saved e g to a memory stick by clicking button Save As 5 5 5 Maintenance...

Page 56: ...art to wash all pipettes needle fill nozzle and rinse nozzle wash station The function can be set up for automatic execution in the following screen Close program will only close down the program 5 5...

Page 57: ...restart sample mode when finished if needed 5 5 8 End of day wash schedule settings Select the time of the day in hours and minutes for automatic start of this function 5 6 Settings screen The Settin...

Page 58: ...1 2 Start pool Ensure that racks are correctly placed in the Start Pool the groove on the side of the rack must catch the strip at the right side of the storage area Note Do not push racks manually a...

Page 59: ...es of the Diluter 4 Prime Nozzle with water activates the fill nozzle water valve De ionized water must flow through the tube connected to the fill nozzle cap 5 Prime Waste system activates the disinf...

Page 60: ...he Interrliner may remain switched ON continuously However the customer should consider environmental issues such as energy consumption when the instrument is not to be used for some time It is also r...

Page 61: ...equence Close the Interrliner software Switch the Interrliner OFF Switch optional printer OFF Switch the ESR analyser unit OFF Switch the Interrliner rack system OFF Operation MRN 164_2 EN Version 2 0...

Page 62: ...U 9 Compact bidirectional 10 Compact unidirectional String format for StaRRsed 11 StaRRsed III V14 12 Vesmatic 13 Sedimatic 15 14 Sedimatic 100 15 Opus bidirectional 16 Advia 120 bidirectional 17 Advi...

Page 63: ...21 60 23 EDTA 905006 5 5 CLEAR 22 60 23 EDTA 079 905007 24 60 23 Too many borders found 905008 25 60 23 L_err 84 75 200 EDTA 905002 905003 905004 Sample results with hazy aspect 905005 Sample result...

Page 64: ...e following codes are defined 0 No errors 1 No cells plasma found Error No contents could be detected in the pipette 2 ESR Probably 140 mm Error Extremely high ESR value 3 Too many borders found Error...

Page 65: ...nute method and temperature correction 75 with a correct column height L_err 84 75 200 means 84 mm in the 60 minute method and temperature correction 75 with a correct column height Note If limit erro...

Page 66: ...see page 131 reports are shown when the change from the hazy level to the cell plasma separation level occurs outside the given distance The following code messages are reported printed in column 5 Th...

Page 67: ...as a standard Starrsed Control should not be used past the expiry date Mechatronics as supplier of the Starrsed Control shall not be liable for any claimed damages arising from other than intended us...

Page 68: ...value and range and the expiry date The history of QC results is maintained internally Error messages are generated when the QC results are outside the acceptable range QC samples can be optionally re...

Page 69: ...es Do not mix residual material with material from other tubes Do not re use empty tubes The software interface is described in the chapter History screen Starrsed Control should be disposed of as med...

Page 70: ...6 2 QC Result analysis Authorized staff should identify and differentiate acceptable unacceptable random errors and trends and or shifts in systematic errors from the statistical data Depending on the...

Page 71: ...must be met 1 Waste tube must not exceed 5 meters or 18 feet in length 2 Drain height must not be higher than the original waste container inside the instrument Disclaimer Check the specifications of...

Page 72: ...entation time 60 or 30 min may be exceeded However the start time is saved and therefore the actual sedimentation time can be checked Important system settings are kept in an internal Flash Eprom insi...

Page 73: ...his section can usually not be solved by the operator Refer to the Service manual for more information available only in English The error numbers are displayed in the PC software 11 1 Errors 183 and...

Page 74: ...bes between the containers and pumps valves Unscrew the cap from the bottle or cubitainer Check the pick up tubes in the bottle or cubitainer Check the tubes for blockages or kinks 11 5 Reagents alarm...

Page 75: ...ontainer blocked Replace the tube Waste pump failure Exchange the waste pump cassette If the error returns call for service Electrical bridge between the waste level electrodes Clean waste separator s...

Page 76: ...growth of micro organisms in the diluter system It is extremely important that the system is kept sterile Erroneous haziness looks more like haemolytic plasma than normal hazy plasma If there is an a...

Page 77: ...ps of Saline to the De ionized water to increase the conductivity 11 11 Air bubbles After a normal aspiration the Westergren pipette must be free of air bubbles In the following examples different pat...

Page 78: ...e fill nozzle washer or O ring 3 Check for air in diluter system 4 Check that the sample probe O ring is not leaking 5 Check transparent mixing piece block for cracks 11 11 2 Pipette looks like zebra...

Page 79: ...op sensor by more than 10mm One air bubble can result in ESR Error 3 11 11 4 One air bubble rising in pipette 1 Usually this is caused by a wet or dirty fill nozzle The blood column should not reach r...

Page 80: ...le tube Small air bubbles result in ESR Error 3 11 11 6 Random air bubbles in pipette 1 Check Diluent flow by priming the diluter system 2 Insufficient sample volume Random air bubbles result in ESR E...

Page 81: ...t of range ESR Result is given Temperature correction not activated Consider QC Sample as not correct Try new QC sample tube normal samples will be finished Check acceptable range in QC settings If th...

Page 82: ...cceptable range The last QC result should be evaluated by authorized staff to decide whether the Interrliner may run patient samples depending on the the nature of errors Press Yes to continue samplin...

Page 83: ...Compare only results from one batch If Lab ID is used check the linked Starrsed Control ID It is possible that a new batch is in use without changing to the new assayed mean value Systematic QC error...

Page 84: ...echanical obstructions and remove these 2 Clear the error by pressing F3 on the keyboard of the Input pool When the error occurs again switch OFF all units and call for service 12 2 Errors 33 and 37 T...

Page 85: ...nstruction Daily WI 257 Daily Weekly WI 258 Weekly maintenance Level 4 maintenance WI 1215 Level 4 maintenance Every 7500 samples Level 3 maintenance WI 224 Level 3 maintenance Every 23000 samples Lev...

Page 86: ...Fill stop sensor box Fill stop sensor FS 90 140 165 Diluter Start sensor check Go to tab Maintenance Check sensor Select Diluter start sensor box Diluter start sensor 400 550 700 Measure sensor check...

Page 87: ...acement New rinse pump tube assembly ESRI090902 New saline pump tube assembly ESRI090903 New tube replacement 1 Open left cover 2 Pull pump tube slightly downwards and at the same time towards the fro...

Page 88: ...ivided in the column Vacuum stabilisation errors may occur 13 3 4 Fill and clean procedure Note Each pipette on the pipette belt will be filled with Starrsed Cleaning Agent after one hour the first pi...

Page 89: ...ed the needle goes back to the home position 4 After the Fill clean process is finished a notification is given to stop the procedure 5 Remove the adapter from the needle assembly and press Stop The i...

Page 90: ...over the sample probe 7 Press Continue 8 The fill and clean process is started 13 4 Level 3 maintenance Level 3 maintenance is level 4 maintenance adding the following extra s 1 Replace the pinch val...

Page 91: ...the sample probe or outer needle Lift the robotic protection hood Needle exchange 1 Unscrew sample probe manually 2 Mark each tube for easier reconnecting to the correct nipple 3 Disconnect the tubes...

Page 92: ...ing door and place the containers on the turntable 2 Remove the cubitainer screw caps and pull the necks of the bottle packs out of the cardboard box 3 Install the level sensors and spacers 4 Add new...

Page 93: ...ter After each reagents change the fluid system must be primed 1 Select Maintenance Prime Clean 2 Perform the applicable prime step to fill the relevant tubes with reagent and remove air 13 7 Cleaning...

Page 94: ...14 Appendix for Interrliner Appendix section Appendix for Interrliner 94 MRN 164_2 EN Version 2 02 Interrliner...

Page 95: ...Sample probe motor is faulty Sample probe motor driver on needle board is faulty Sample probe is blocked E5 Duplicated ID Sample rejected Sample already in carousel Wait until sample is measured Check...

Page 96: ...nozzle motor is faulty Fill nozzle motor driver on nozzle board is faulty Fill nozzle is blocked E14 Outer needle motor position error home Outer needle did not reach the home top sensor within a cer...

Page 97: ...setting Selected result path is not valid Software is using the default setting Check result path setting Check if network or USB devices are used E30 No ACK NACK received from host after sending inqu...

Page 98: ...osition Could not start the position motor because the outer needle or sample probe is not its home position top Check outer needle home sensor Check sample probe home sensor Faulty Outer needle motor...

Page 99: ...sum error Indexer settings After reading the settings from EEPROM the calculated checksum did not match with the checksum stored in EEPROM Default Indexer settings are loaded Faulty EEPROM 24C01 on El...

Page 100: ...ted Bad I2C cable s Note the boards in the ESR analyser unit are connected in series E169 I2C communication error between PC and Drive board There was a communication error between the Driver board an...

Page 101: ...n motor Check sensors of Barcode rotator up down unit E185 Indexer error elevator platform not in back position The indexer could not move to a new position because the elevator platform was not in it...

Page 102: ...tform slider unit could not move to a new position because the sample arm was not in its out position rack position Check rotation motor Check out sensor of Sample arm E200 Sample unit in out and Tube...

Page 103: ...k the InteRRliner I2C cable s E213 Bad communication between ESR unit and Input pool The ESR analyser unit did not receive a good answer second part after sending an inquiry to the Input pool Check RS...

Page 104: ...End of day wash X X X X X X Clean outside aspiration needle X X X X X X Check tubing diluent syringe X X X X X X Clean outside instrument X X X X X X Clean Fill nozzle X X X X X Clean waste separator...

Page 105: ...Replace Sample probe assembly X ESRI050909 1 Replace Pipette valves bodies 84 pieces X QTST040001 84 Replace Outer needle assembly X ESRI050950 1 Replace all tubing with Tubing set X EHST079200 1 Com...

Page 106: ...SRI070924 Fill tube assembly to regulator ESRI070924 Fill tube assembly from regulator QWLV030021 Diluter Syringe QWLV080002 Quad Ring Appendix for Interrliner 106 MRN 164_2 EN Version 2 02 Interrline...

Page 107: ...15 Work instruction Interrliner Work instruction section Work instruction Interrliner MRN 164_2 EN Version 2 02 Interrliner 107...

Page 108: ...bes and connections for leaks Check that liquid does not run back after the pumps have stopped 3 Clean the outer needle with disinfectant 4 Check tubing from the syringe for trapped air bubbles 5 Chec...

Page 109: ...liner Revision 001 2019 Clean Fill nozzle 1 2 3 4 Clean fill nozzle The use of a toothbrush and detergent is recommended Carefully scrub the fill nozzle inner part Use a tissue to dry the fill nozzle...

Page 110: ...read of the glass jar the top rim of the glass jar the O ring for the HEPA filter and the black PVC pipe on the back Replacing C If applicable replace the bacterial HEPA filter For Maintenance Level 4...

Page 111: ...nance Check sensor Select Check Diluent flow sensor box Press test When test is finished signal Down and signal Up must be green Separator check Go to tab Maintenance Check sensor Select Check Separat...

Page 112: ...Prime Clean and perform the Prime Diluent Diluter function 7 Wipe outer surface and stainless steel plate below the pipettes with disinfectant 8 Check or replace if needed two drip trays ESRI010318 W...

Page 113: ...ner Revision 003 2019 Clean Fill nozzle and exchange O ring Fill Nozzle 1 2 3 4 5 6 Clean fill nozzle The use of a toothbrush and detergent is recommended Carefully scrub the fill nozzle inner part Us...

Page 114: ...all parts with hot water and a brush B Use some acid free vaseline on the screw thread of the glass jar the top rim of the glass jar the O ring for the HEPA filter and the black PVC pipe on the back...

Page 115: ...p tube slightly downwards and at the same time towards the front of the unit to release the tube out of the pump plate holder 3 Remove the old tube from the peristaltic pump rotor 4 Disconnect the tub...

Page 116: ...l 4 Put the adapter with cleaning solution on the lower tube holder 5 Select Maintenance tab Prime Clean button Fill and clean Start Fill and clean procedure 1 Select button OK 2 The needle goes down...

Page 117: ...sor check Go to tab Maintenance Check sensor Select Check Diluent flow sensor box Press test When test is finished signal Down and signal Up must be green Separator check Go to tab Maintenance Check s...

Page 118: ...Prime Clean and perform the Prime Diluent Diluter function 7 Wipe outer surface and stainless steel plate below the pipettes with disinfectant 8 Check or replace if needed two drip trays ESRI010318 W...

Page 119: ...ner Revision 004 2019 Clean Fill nozzle and exchange O ring Fill Nozzle 1 2 3 4 5 6 Clean fill nozzle The use of a toothbrush and detergent is recommended Carefully scrub the fill nozzle inner part Us...

Page 120: ...ssembly ESRI090903 New tube replacement 1 Open left cover 2 Pull pump tube slightly downwards and at the same time towards the front of the unit to release the tube out of the pump plate holder 3 Remo...

Page 121: ...it Fill and clean This cycle takes about 90 minutes 1 Fill the clean adapter EHST110907 with hot de ionized water 150 ml 80 C 2 Add 15 ml Starrsed Cleaning Agent QRR 010905 to the hot water in the ada...

Page 122: ...Check sensor Select Check Fill stop sensor box Fill stop sensor FS 90 140 165 Diluter Start sensor check Go to tab Maintenance Check sensor Select Diluter start sensor box Diluter start sensor 400 550...

Page 123: ...ube with use of the lever 3 Pull the waste separator towards the front of the ESR analyser unit Note The separator has two sensor connectors at the rear 4 Remove bacterial HEPA filter 5 Fill waste sep...

Page 124: ...place 7 Remove the protection caps on from the tubes 8 Connect the two tubes from the waste pump cassette on the luers at the steel plate under the waste pump Exchange Pinch valve tube Replace the pin...

Page 125: ...trapped air bubbles 6 If trapped air bubbles are found go to tab Maintenance click button Prime Clean and perform the Prime Diluent Diluter function 7 Wipe outer surface and stainless steel plate bel...

Page 126: ...disinfectant 6 Leave the disinfectant in the system for 15 minutes Step 2 1 Take the diluent suction tube out of the disinfectant 2 Wipe the tube clean and dry with a tissue 3 Empty the bottle and ref...

Page 127: ...l parts with hot water and a brush B Use some acid free vaseline on the screw thread of the glass jar the top rim of the glass jar the O ring for the HEPA filter and the black PVC pipe on the back Rep...

Page 128: ...s cycle takes about 90 minutes 1 Fill the clean adapter EHST110907 with hot de ionized water 150 ml 80 C 2 Add 15 ml Starrsed Cleaning Agent QRR 010905 to the hot water in the adapter 3 Place the cap...

Page 129: ...e needle goes back to the home position 4 After the Fill clean process is finished a notification is given to stop the procedure 5 Remove the adapter from the needle assembly and press Stop The instru...

Page 130: ...edle 4 Pull the sample probe complete with outer needle towards the front edge of the ESR analyser unit The outer needle must be supported to protect it from falling 5 Slide the new sample probe into...

Page 131: ...of sedimentation setting of erythrocytes red blood cells in a blood column during a specified time Hazy A sedimentation is reported to be hazy when the boundary between blood plasma and erythrocytes...

Page 132: ...f day wash procedure 60 52 End of day wash schedule settings 57 EQAS 12 Errors 165 171 and 210 214 73 Errors 181 185 and 193 201 74 Errors 183 and 193 73 Errors 24 26 38 42 43 62 67 70 75 80 84 Errors...

Page 133: ...rror messages 69 QC Normal results table 37 QC Result analysis 70 42 QC Results 69 QC Results screens 36 Quality control 67 Quality control procedure 68 Quality control troubleshooting 80 Quick start...

Page 134: ...U Usage options 68 Used reagents 19 W Waste disposal 71 Weekly 86 Work instruction Interrliner 107 134 MRN 164_2 EN Version 2 02 Interrliner...

Reviews: