2-5
2.3
INSTALLING THE SYSTEM
Place the system on a flat, level, stable surface that positions the system at a
comfortable height for the operator. Connect the power cord from the rear panel to a
properly grounded power supply outlet.
Rear Panel Connections
2.3.1 Supplies
ELECTRICAL 110-120VAC, 2A, 50-60 Hz, single phase
FUSE 2A F
NON- ACCESSABLE FUSES (P/S within enclosure)
GPC40-A
2A F
GPC41-24
2A F
2.3.2 Environmental Requirements
The System 610 is intended to be installed in a facility meeting the following
environmental requirements unless otherwise specified in a particular section.
i.
Indoor use
ii.
Altitude up to 2000 meters
iii.
60 to 80 degrees Fahrenheit
iv.
Relative humidity of 10 to 90 percent, non-saturating.
v.
Main supply voltage fluctuations not to exceed ±10% of nominal
voltage.
vi.
Transient over-voltages according to installation categories:
Overvoltage Categories II
vii. Pollution Degree 2
viii. Positioning:
The System must be placed on a leveled surface
(fluid level (sensitivity .035 in/ft) included).
If the equipment is used in a manner not specified by the
manufacturer, the protection provided by the equipment may
be impaired.
Summary of Contents for 610
Page 2: ......