I
NITIAL
S
TARTUP AND
C
ALIBRATION
748460-A
3-21
Rosemount Analytical
January 2000
3.14.1 STATUS D
ISPLAY
Refer to Figure 3-13. The STATUS display shows the alarms, error messages and
security lockout status. (See explanations of error messages in Section 5.1.) The
STATUS display can be used to check the following alarm setpoints without entering
one of the MODE functions: HIGH/LOW, AUTO/MANUAL and ON/OFF.
The order of priority for error messages, security status and alarms is as follows:
[SECURITY ENABLED/DISABLED]
[ERL ]
[EL-LIN.COEFF ERR]
[E0-ZERO POT LMTS]
[E1-R1 SP POT LMTS]
[E2-R2 SP POT LMTS]
[E3-R3 SP POT LMTS]
[E4-ADC SATURATED]
[E5-ZERO DRIFT]
[E6-SPAN DRIFT]
[E7-GAIN TOO HIGH]
[RMT: R/L]
[ALARM 1 AUTO/{MAN ON/OFF}]
[ALARM 2 AUTO/{MAN ON/OFF}]
[CAL-CTL PRESENT]
[AUTOCAL: ON/OFF]
[CURRENT 0/4 SP ON/OFF]
F
IGURE
3-13. S
TATUS
D
ISPLAY
Alarm:
MN (Manual) or
AT (Automatic)
Alarm 1 (15 sec.)
{Followed by]
Alarm 2 (15 sec.)
Alarm ON
or OFF
ALARM 1 MN OFF
Summary of Contents for 890
Page 1: ...Rosemount Analytical MODEL 890 UV ANALYZER INSTRUCTION MANUAL 748460 A ...
Page 8: ...CONTENTS VI 748460 A Rosemount Analytical January 2000 NOTES ...
Page 16: ...PREFACE Rosemount Analytical January 2000 748460 A P 8 NOTES ...
Page 64: ...TROUBLESHOOTING Rosemount Analytical 748460 A January 2000 5 4 NOTES ...
Page 84: ......
Page 85: ......
Page 86: ......
Page 87: ......
Page 88: ......
Page 89: ......
Page 90: ......
Page 91: ......
Page 92: ......
Page 93: ......
Page 94: ......
Page 95: ......
Page 96: ......