Rosemount Analytical
748460-A
January 2000
3-16
0 500 1000 1500 2000 2500 3000 3500 4000 4500 5000
0
CONCENTRATION
ppm
DISPLAY
ppm
Display ppm
500
1000
2000
2500
3000
3500
4000
4500
5000
1500
Linearization coefficients can be developed for each range and stored in the analyzer
using the front panel keypad. The operating range is selected by entering RANGE = 1,
2, or 3 in the Range Mode. Coefficients for the selected range are automatically used,
independent of the position of Range = n, from the Linearization Mode (except for the
ALL Range selection).
FULLSCALE
A0
A1
A2
A3
A4
50 ppm
0.00016
0.96898
0.06899
-0.08882
0.05043
100 ppm
-0.00015
0.95033
0.07541
-0.01677
-0.00861
250 ppm
0.00014
0.92430
0.05284
0.06604
-0.04362
500 ppm
0.00023
0.87214
0.03032
0.20556
-0.10857
1000 ppm
0.00012
0.76590
0.11188
0.18932
-0.06724
2500 ppm
0.00003
0.53242
0.31440
-0.12316
0.27611
5000 ppm
0.00071
0.00859
2.28880
-1.02406
2.71778
Consult factory for coefficients for Cl
2
ranges.
T
ABLE
3-1. L
INEARIZATION
C
OEFFICIENTS
, S
TANDARD
SO
2
R
ANGES
F
IGURE
3-9. T
YPICAL
A
PPLICATION
L
INEARIZATION
C
URVE
Summary of Contents for 890
Page 1: ...Rosemount Analytical MODEL 890 UV ANALYZER INSTRUCTION MANUAL 748460 A ...
Page 8: ...CONTENTS VI 748460 A Rosemount Analytical January 2000 NOTES ...
Page 16: ...PREFACE Rosemount Analytical January 2000 748460 A P 8 NOTES ...
Page 64: ...TROUBLESHOOTING Rosemount Analytical 748460 A January 2000 5 4 NOTES ...
Page 84: ......
Page 85: ......
Page 86: ......
Page 87: ......
Page 88: ......
Page 89: ......
Page 90: ......
Page 91: ......
Page 92: ......
Page 93: ......
Page 94: ......
Page 95: ......
Page 96: ......