Reference Manual
00809-0100-4026, Rev FA
July 2009
Rosemount 5400 Series
C-12
Surface Echo Tracking
The Surface Echo Tracking function can eliminate ghost echo problems below
the product surface. This may occur in Still-pipes because of multiple
reflections between the pipe wall, flange and antenna. In the tank spectrum
these echoes appear as amplitude peaks at various distances below the
product surface.
To activate this function, select the
Always Track First Echo
check-box making
sure there are no disturbing echoes above the product surface when this
function is activated.
Amplitude
Dist
ance
First echo corresponding
to the product surface
Surface