Accuracy
Reference Accuracy Example (HART only)
When using a Pt 100 (
= 0.00385) sensor input with a 0 to 100 °C span:
Digital Accuracy would be ±0.10 °C, D/A accuracy would be ±0.02% of 100 °C
or ±0.02 °C, Total = ±0.12 °C.
Differential Capability Exists Between Any Two Sensor Types
(dual-sensor option)
For all differential configurations, the input range is X to Y where:
• X = Sensor 1 minimum – Sensor 2 maximum
and
• Y = Sensor 1 maximum – Sensor 2 minimum.
Sensor Options
Sensor Reference
Input Ranges
Minimum
Span
(1)
Digital
Accuracy
(2)
D/A
Accuracy
(3)(4)
2-, 3-, 4-wire RTDs
°C
°F
°C
°F
°C
°F
Pt 100 (
= 0.00385)
IEC 751
–200 to 850
–328 to 1562
10
18
± 0.10
± 0.18
±0.02% of span
Pt 200 (
= 0.00385)
IEC 751
–200 to 850
–328 to 1562
10
18
± 0.22
± 0.40
±0.02% of span
Pt 500 (
= 0.00385)
IEC 751
–200 to 850
–328 to 1562
10
18
± 0.14
± 0.25
±0.02% of span
Pt 1000 (
= 0.00385)
IEC 751
–200 to 300
–328 to 572
10
18
± 0.10
± 0.18
±0.02% of span
Pt 100 (
= 0.003916)
JIS 1604
–200 to 645
–328 to 1193
10
18
± 0.10
± 0.18
±0.02% of span
Pt 200 (
= 0.003916)
JIS 1604
–200 to 645
–328 to 1193
10
18
± 0.22
± 0.40
±0.02% of span
Ni 120
Edison Curve No. 7
–70 to 300
–94 to 572
10
18
± 0.08
± 0.14
±0.02% of span
Cu 10
Edison Copper Winding No. 15
–50 to 250
–58 to 482
10
18
±1.00
± 1.80
±0.02% of span
Pt 50 (
=0.00391)
GOST 6651-94
–200 to 550
–328 to 1022
10
18
±0.20
±0.36
±0.02% of span
Pt 100 (
=0.00391)
GOST 6651-94
–200 to 550
–328 to 1022
10
18
±0.10
±0.18
±0.02% of span
Cu 50 (
=0.00426)
GOST 6651-94
–50 to 200
–58 to 392
10
18
±0.34
±0.61
±0.02% of span
Cu 50 (
=0.00428)
GOST 6651-94
–185 to 200
–301 to 392
10
18
±0.34
±0.61
±0.02% of span
Cu 100 (
=0.00426)
GOST 6651-94
–50 to 200
–58 to 392
10
18
±0.17
±0.31
±0.02% of span
Cu 100 (
=0.00428)
GOST 6651-94
–185 to 200
–301 to 392
10
18
±0.17
±0.31
±0.02% of span
Thermocouples
(5)
Type B
(6)
NIST Monograph 175, IEC 584
100 to 1820
212 to 3308
25
45
± 0.75
± 1.35
±0.02% of span
Type E
NIST Monograph 175, IEC 584
–50 to 1000
–58 to 1832
25
45
± 0.20
± 0.36
±0.02% of span
Type J
NIST Monograph 175, IEC 584
–180 to 760
–292 to 1400
25
45
± 0.25
± 0.45
±0.02% of span
Type K
(7)
NIST Monograph 175, IEC 584
–180 to 1372
–292 to 2501
25
45
± 0.25
± 0.45
±0.02% of span
Type N
NIST Monograph 175, IEC 584
–200 to 1300
–328 to 2372
25
45
± 0.40
± 0.72
±0.02% of span
Type R
NIST Monograph 175, IEC 584
0 to 1768
32 to 3214
25
45
± 0.60
± 1.08
±0.02% of span
Type S
NIST Monograph 175, IEC 584
0 to 1768
32 to 3214
25
45
± 0.50
± 0.90
±0.02% of span
Type T
NIST Monograph 175, IEC 584
–200 to 400
–328 to 752
25
45
± 0.25
± 0.45
±0.02% of span
DIN Type L
DIN 43710
–200 to 900
–328 to 1652
25
45
± 0.35
± 0.63
±0.02% of span
DIN Type U
DIN 43710
–200 to 600
–328 to 1112
25
45
± 0.35
± 0.63
±0.02% of span
Type W5Re/W26Re
ASTM E 988-96
0 to 2000
32 to 3632
25
45
± 0.70
± 1.26
±0.02% of span
GOST Type L
GOST R 8.585-2001
–200 to 800
–328 to 1472
25
45
± 0.25
± 0.45
±0.02% of span
Other Input Types
Millivolt Input
–10 to 100 mV
3 mV
±0.015 mV
±0.02% of span
2-, 3-, 4-wire Ohm Input
0 to 2000 ohms
20 ohm
±0.35 ohm
±0.02% of span
(1) No minimum or maximum span restrictions within the input ranges. Recommended minimum span will hold noise within accuracy specification with damping at
zero seconds.
(2) Digital accuracy: Digital output can be accessed by the Field Communicator.
(3) Total Analog accuracy is the sum of digital and D/A accuracies.
(4) Applies to HART / 4-20 mA devices.
(5) Total digital accuracy for thermocouple measurement: sum of digital ac0.25 °C (0.45 °F) (cold junction accuracy).
(6) Digital accuracy for NIST Type B is ±3.0 °C (±5.4 °F) from 100 to 300 °C (212 to 572 °F).
(7) Digital accuracy for NIST Type K is ±0.50 °C (±0.9 °F) from –180 to –90 °C (–292 to –130 °F).