Power Supply
Item
Description
Power supply
Receives power from the Accura 2300[S] device via the RJ12 communication cable
Current consumption
145 mA current is consumed when the backlight of the LCD screen and all LEDs are turned
on, and digital outputs are activated.
1
1. When the Accura 2350-TEMP module is connected with the Accura TSEN device, 15 mA current is additionally
consumed per Accura TSEN device.
LCD Screen
Module
LCD
Accura 2350-TEMP
2.7
”
Graphic 80x128 Dots (40.0 x 58.0mm) FSTN LCD
The LCD screen consists of three operational modes as follows.
Mode
Description
DISPLAY mode
Displays measured values
SETUP mode
Enables to check setup items and change it
EVENT mode
Displays detailed event data
LED Status Indicators
Item
Color
Description
POWER
Green
It is turned on when the device is powered on.
EVENT
Red
It blinks red when an event occurs.
MODULE
Orange
It blinks orange when the Accura 2350-TEMP module normally communicates
with Accura 2300[S].
TSEN
Orange
It blinks orange when the Accura 2350-TEMP module normally communicates
with Accura TSEN.
LCD Control Button
Item
Operation
Description
Left[<]
Short press
1
Used to move to the left
Up[
∧
]
Short press
Used to move upward
Down[
∨
]
Short press
Used to move downward
Right[>]
Short press
Used to move to the right
ESC
Long press
2
Used to perform the Escape function
SETUP
Long press
Used to change the display mode to the SETUP mode
EVENT
Long press
Used to change the display mode to the EVENT mode
ENTER
Long press
Used to perform the Enter function
1. Press the button and release it immediately.
2. Release the button after pressing it for 0.5 seconds.
Summary of Contents for Accura 2350-TEMP
Page 1: ......