NEXUS CALL SYSTEM
by Rondish
P31
2.5 DoorWatcher
DoorWatcher monitors offer protection against wandering from at-risk patients, especially useful for
facilities with memory care residents. The monitors provide a visual alarm indication when a resident tag
moves within range, can activate a door lock for delayed egress, and signal event details to Nexus for a
caregiver to respond.
Refer to DoorWatcher User Instructions for further details of how this equipment operates.
2.5.1 DoorWatcher Settings
To use a DoorWatcher sensor with Nexus, the sensor strip should first be programmed, and it can then be
added to Nexus in Device Management (Section 3.4). Remove the end cap of a DMS-21 sensor and pull
out the PCB to access switches and to program the sensor
Switches
–
Basic Setup
Volume: Set switch to Hi/Lo/Off
Detection distance: Set to Short/Med/Long
Extension: Set as Primary/Secondary
Night mode:
Programming
–
Advanced Setup
Enter programing by holding the magnetic button for 3s. Select the value of each parameter using the
rotary dial, and trigger the magnetic button to enter the programming value.
In total you should be
inputting five digits, at which point the DMS-21 will automatically exit programming.
Door ID
(two digits) (01 default)
Site code
: 0-9 (0 default)
Lock timer
–
this will control how long a door lock is active to delay or prevent egress through a protected
door.
Setting
Lock Time
0
Indefinite (Default)
1
10s
2
15s
3
20s
4
25s
5
30s
6
35s
7
40s