Copyright ROMPA® Ltd
GYMTOP® USB PROFESSIONAL 20143, KH 2.10
The button
Choose
switches in/out of the
Patient selection
and the pertinent exercises if
necessary. This way you can look at the valuation of the exercises on a full screen. You can
highlight the patients or the exercises with a click of the mouse. In the evaluation, the
appropriate exercise is then marked by a red point. The diagram shows the exercises
chronologically (first at top; last at the bottom).
As more evaluations are added, trends and patterns can be monitored. Evaluations can be
highlighted and deleted by pressing the delete button.
SAFETY
1.
Always supervise the use of this product.
CARE AND MAINTENANCE
1.
Ambient conditions: for transportation and storage, keep this product at a temperature
between -10°C to +50°C.
2.
Operating conditions: keep this product at a temperature b10°C and +40°C;
relative humidity 30-75%; atmospheric pressure 700-1060 hPa.
3.
Once this product is eventually exhausted, please dispose of it appropriately and
responsibly.
SPECIFICATION
Size: 9 x 4 x 2.5cm
Weight: 90g
Socket: USB-Plug type A for connection to a PC
Power supply: USB-port of PC 35mA and 5V
Security options: Protection class II; IP 20
EMV compatibility: Emission: EN 55011 Class A
EN 55014 for click noise
Immunity: ESD (IEC 61000-4-2)
2 + 4 + 6kV Contakt, 2 + 4 + 8kV Air
Radiated field (IEC 61000-4-3); 3V/m, 80-2500 MHz, AM: 80%, 1kHz or 2 Hz, Modulation
Frequency; 10V/m from 80-2500MHz
Burst: (IEC 61000-4-4) 2 kV
Surge: (IEC 61000-4-5)
0.5kV, 1kV differential mode
0.5kV, 2kV common mode
Conducted RF Immunity (IEC 61000-4-6)
3V, 80% AM, 0.15-80 MHz, 1 kHz or 2Hz Modulation Frequency; different startfreq for battery
powered systems
Magnetic Fields (IEC 61000-4-8) 50 Hz, 10A/m
AC variations (IEC 61000-4-11)
Voltage Test Level 0%, 40 and 70%
Duration (Periods) 0.5, 5, 25; Voltage Interruption 5 sec