(IEC 60601-1-2)
Table 4: Guidance and
Manufacture’s Declaration for Electromagnetic
Immunity for non-life support systems
The
RomiApex™ A-15 is intended for use in the electromagnetic environment specified below. The
customer or user should ensure that it is used in such an environment.
IMMUNITY
test
IEC 60601 test
level
Compliance level
Electromagnetic environment
guidance
Conduction RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms
150 kHz to 80 MHz
3 V/m
80 MHz to 2.5 GHz
3 Vrms
150 kHz to 80 MHz
3 V/m
80 MHz to 2.5 GHz
RomiApex™ A-15 is not affected
by portable and mobile RF
communications equipment in
close proximity.
There is no recommended
separation distance from
transmission sources.
RomiApex™ A-15 will operate
safely and effectively near cell
towers and radio/TV transmitters.
Note 1: The
RomiApex™ A-15 may be susceptible strong magnetic fields or static electric fields
higher than the limits specified in testing, which could interrupt programming. If there is a
reason to believe that this type of interference has occurred, simply re-insert the battery
and the unit should reset.