5.4.14.  The  file  clip  may  be  disconnected  from  the  file  and  connected  back  during 

measurement without affecting normal device  operation (for instance, when 
the  file  is  changed  to  a  larger  number  during  root  canal  treatment  or  when 
another  canal  should  be  measured).  In  such  cases  the  device  detects 
automatically that the new measurement is initiated, the electrical contact and 
conductivity of the root canal are checked again and two beeps are sounded. 

 

Note:

  The 

RomiApex™  A-15  operates  completely  automatically.  No  manual 

adjustments  are  required.  The 

RomiApex™  A-15  enables  accurate  localization  of 

apical foramen independently of root canal conditions (dry, wet, with blood, pulp). In 
case of very dry canal or previous obturation (retreatment cases) you may use irrigation 
solution such as Sodium Hypochlorite, Saline, etc. to provide a conductive electrical 
environment. 

 

5.5. 

Audio feedback 

The 

RomiApex™ A-15

 

is equipped  with  an audio  indicator which  is  activated  in  parallel 

with progression of the file. This function enables monitoring of the file progression within 
the canal in the apical zone even without seeing the display. 

The  volume  can  be  adjusted  to  one  of  the  four  levels: mute,  low,  normal  and  high.  The 

adjustment is performed by successive pressings of the volume key 

5.6. 

Automatic Shutdown 

The 

RomiApex™  A-15

 

automatically  shuts  off  after  5  minutes  without  use.  In  order  to 

prolong the battery life, after completing the measurements, it is recommend switching the 

device off by pressing the On/Off key

   

 

 

Summary of Contents for RomiApex A-15

Page 1: ...RomiApex A 15 APEX LOCATOR User Manual Please scan the QR code to view RomiApex A 15 additional instructions...

Page 2: ...y 7 5 3 Cable connection test 8 5 4 Device operation 9 5 5 Audio feedback 12 5 6 Automatic Shutdown 12 6 Demo mode 13 7 Maintenance cleaning and sterilization 13 8 Troubleshooting guide 15 9 Certifica...

Page 3: ...ietary algorithms to determine the position of the apical foramen Utilizing the advantages of its well proven patented technology RomiApex A 15 offers the operator performing root canal treatments an...

Page 4: ...nit with patients who have a pacemaker as its effect has not been evaluated RomiApex A 15 should not be used in the presence of flammable anesthetic mixtures with air or with oxygen or nitrous oxide R...

Page 5: ...tions Battery cannot be disposed of as municipal waste within the European Union Ensure the battery is installed with correct polarity Do not immerse RomiApex A 15 unit This may damage the device and...

Page 6: ...the package before use 1 RomiApex A 15 1 Cradle 1 AAA 1 5V Alkaline Battery 1 Measuring cable 2 File clips 5 Lip clips 1 Touch probe 1 Screwdriver for battery compartment 1 User Manual 1 Measuring ca...

Page 7: ...tenance of the RomiApex A 15 5 2 Installing Replacing the Battery The RomiApex A 15 is powered by a standard AAA size 1 5V alkaline battery included 5 2 1 To install replace the battery unscrew and re...

Page 8: ...h use the RomiApex A 15 should be checked for functionality The RomiApex A 15 has a connection test feature in order to check the cables 5 3 1 Connect the measuring cable to the device Connect the fil...

Page 9: ...t a 15 hand K file or other appropriate file into the canal 5 4 6 Press the On Off button to turn the device on The initial image appears on the display 5 4 7 Plug the measuring cable into the device...

Page 10: ...s toward the foramen color segments inside the root canal image accompanied by audio signals with varied frequency will indicate the file s progress Numerical readings 2 0 1 75 1 5 1 25 1 0 0 75 0 5 0...

Page 11: ...electronically obtained length using an endodontic ruler To establish the working length subtract 0 5 1 0mm from the electronic length The red OVER indication accompanied by frequent beeps indicates t...

Page 12: ...dry wet with blood pulp In case of very dry canal or previous obturation retreatment cases you may use irrigation solution such as Sodium Hypochlorite Saline etc to provide a conductive electrical en...

Page 13: ...mo cycle RomiApex A 15 exits Demo mode and switches automatically to regular operation mode 7 Maintenance Cleaning and Sterilization 7 1 General The device does not contain user serviceable parts The...

Page 14: ...ak the accessories in a disinfectant solution that contains a proteolytic enzyme if possible Rinse accessories thoroughly in running water Follow manufacturer s instructions on the disinfectant concen...

Page 15: ...re lip clip is contacting the oral mucosa and file clip is clean and free of debris Fill the canal with an electrolytic solution if required i e Sodium Hypochlorite Saline etc The endodontic file is n...

Page 16: ...lip dirty Clean the file clip with Ethanol 70 80 vol Is blood or other fluids over flowing the access cavity of the tooth If blood or other fluids are overflowing the access cavity they may create a c...

Page 17: ...surfaces infected with caries Deep caries may create a conductive path outside the canal and cause incorrect measurements OVER indication unstable readings etc Are there external resorption or is the...

Page 18: ...54 Fax 32 2 732 60 03 E Mail mail obelis net 11 Limited Warranty RomiApex A 15 is warranted for 24 months from the date of purchase The accessories Measuring cable file clips lip clips touch probe are...

Page 19: ...ut obligation to notify any person of such changes modifications or revisions 13 Technical Specifications The RomiApex A 15 belongs to the following category of medical devices Internally powered equi...

Page 20: ...sions EN 5011 European Norm equivalent to CISPR 11 Group 1 The RomiApex A 15 uses low voltage and low current electrical and electromagnetic energy only for its internal functions Therefore its RF emi...

Page 21: ...ine to line 2 kV line to earth Not applicable Not applicable Voltage dips shorts interruptions and variations on the power supply input lines IEC 61000 4 11 5 U 95 dip in U for 0 5 cycle 40 U 60 dip i...

Page 22: ...80 MHz 3 V m 80 MHz to 2 5 GHz 3 Vrms 150 kHz to 80 MHz 3 V m 80 MHz to 2 5 GHz RomiApex A 15 is not affected by portable and mobile RF communications equipment in close proximity There is no recomme...

Page 23: ...nsmitter W Separation distance according to frequency of transmitter m 150 kHz to 80 MHz d 0 35 P 80 MHz to 800 MHz d 0 35 P 800 MHz to 2 5 GHz d 0 7 P 0 01 0 04 0 04 0 07 0 1 0 11 0 11 0 22 1 0 35 0...

Page 24: ...s Special disposal of waste electrical and electronic equipment Follow instructions for use Direct current Type BF applied part Manufacturer SN Serial Number Caution Federal law restricts this device...

Page 25: ...RomiApex A15 Rev 07 04 14 Manufacturer Romidan Ltd 5 Simcha Holzberg St 55022 Kiryat Ono Israel Email export romidan com www romidan com...

Reviews: