20
14. Electromagnetic declarations
(IEC 60601-1-2)
Table 1:
Guidance and
Manufacture’s Declaration for Electromagnetic
Emissions
The
RomiApex™ A-15 is intended for use in the electromagnetic environment specified below.
The customer or user should ensure that it is used in such an environment.
Emissions Test
Compliance
Electromagnetic environment -
guidance
RF emissions
EN 5011
(European Norm equivalent to CISPR
11)
Group 1
The
RomiApex™ A-15 uses low voltage
and low current, electrical and
electromagnetic energy only for its internal
functions. Therefore, its RF emissions are
very low and are not likely to cause
interference in nearby electronic
equipment.
RF emission
EN 5011
(European Norm equivalent to CISPR
11)
Class B
Harmonic emissions
IEC 61000-3-2
NA
The
RomiApex™ A-15 does not have AC
power input or AC power output, therefore
testing for this is not applicable.
Voltage fluctuations/flicker emissions
IEC 61000-3-3