7.2.
Disinfection and sterilization procedure
#
Operation
Instructions
Details and Warnings
1
Preparation at the
point of use prior
to processing
No particular
requirements.
2
Preparation for
decontamination/
preparation before
cleaning
No particular
requirements.
3
Cleaning:
Automated
The accessories are
not intended for
automated cleaning.
4
Cleaning: Manual
Clean the accessories
with an appropriate
brush or towel soaked
in a disinfectant
solution.
- The file clip should be pressed
and released several times
during cleaning to assure all
debris is removed.
- After cleaning, there should be
no visible debris on the
accessories.
5
Disinfection
Soak the accessories
in a disinfectant
solution that contains
a proteolytic enzyme if
possible. Rinse
accessories
thoroughly in running
water.
-
Follow manufacturer’s
instructions on the disinfectant
(concentration, immersion
time, etc.).
-
Do not use disinfectant
containing aldehyde, phenol
or any products which may
damage the items.
6
Drying
No particular
requirements.
7
Maintenance,
inspection and
testing of the
accessories
No particular
requirements.
8
Packaging
Pack the devices in
sterilization pouches.
- Check the expiration date of
the pouch given by the
manufacturer to determine the
shelf life.
- Use packaging which is
resistant up to a temperature of
141°C (286°F).
9
Sterilization
- Steam sterilization at
135°C (275°F) for 10
minutes in table top,
N-type autoclave.
- Follow maintenance and
operation procedures of the
autoclave provided by the
manufacturer.