Configuring the RF signal
R&S
®
SMM100A
313
User Manual 1179.1341.02 ─ 05
"External"
Uses an external reference signal.
Note:
If the external reference is missing, the R&S
SMM100A issues
a warning message and indicates the icon (external reference
missing).
To set the frequency of the external reference, see
Remote command:
Show Connector
Accesses a dialog that displays the physical location of the selected connector on the
front/rear panel of the instrument.
A blinking LED on the front/rear panel indicates the selected connector, too (see
ter 11.2.2.4, "Status indicators"
Deactivate RF Output (if external reference is missing)
Turns off the RF output when the external reference signal is selected, but no signal is
supplied.
This function prevents that no improper RF signal due to the missing external refer-
ence signal is used for measurements. A message indicates that the RF output is
deactivated.
Remote command:
[:SOURce]:ROSCillator:EXTernal:RFOFf[:STATe]
External Reference Frequency
Selects the frequency of the external reference signal.
Set the reference output frequency with the parameters in the
"10
MHz"
Selects 10
MHz for the external reference frequency signal.
"100
MHz, 1
GHz"
Option: R&S
SMM-K703
Selects the external reference frequency.
"Variable"
Option: R&S
SMM-K704
The external reference signal has an arbitrary frequency, within the
permissible range from 1
MHz to 100
MHz.
Remote command:
[:SOURce]:ROSCillator:EXTernal:FREQuency
Variable Reference Frequency
Option: R&S
SMM-K704
Sets the variably settable external reference frequency.
Remote command:
[:SOURce]:ROSCillator:EXTernal:FREQuency:VARiable
Reference oscillator