System configuration
R&S
®
SMBVB-K362, R&S
®
SMW-K362
11
User Manual 1178.9126.02 ─ 04
c) In the tab "Automation", configure the DUT-related automation methods, that
are used for DUTs commands.
See also section DUT Files and Automation of the R&S
CMWrun base manual.
The test module supports the AT commands via COM ports and ADB com-
mands. You can also mix control methods, e.g., ADB shell command sending
with COM port NMEA reading. The default port is specified via "Resources" >
"Serial Port".
Set all commands for which parameters are available. Furthermore, it is impor-
tant to set the control method for the "IVS" -> "Read Back" function to read the
NMEA stream properly.
For AT commands, map the serial ports accordingly to bi-directional or two-way
usage.
● Bidirectional: uses common serial port for IVS configuration and the trans-
mission of NMEA messages.
● Two-way: configurates two serial ports to separate control connection and
the connection for NMEA data.
Preparation of IVS connection