P r o b e R T- Z S 1 0 L
General
Operating Basics
When using this active probe device make sure
the measuring instrument is set to 50 Ω input
coupling and the probe is connected to the
power supply.
In order to get best measurement results
please make sure that the power cord has the
maximum possible distance to the probe tip.
Handling
Note that the probe cable is a sensitive part of
the probe. Do not damage through excessive
bending or pulling. Avoid mechanical shock to
this product in general to guarantee accurate
performance and protection.
Cleaning
To clean the exterior of the probe use a soft
cloth moistened with either distillated water or
isopropyl alcohol. Before use allow the probe to
dry completely.
Warranty and repair
Our
instruments are subject to strict quality con
-
trols. Prior to leaving the manufacturing site, each
instrument undergoes a 10-hour burn-in test. This
is followed by extensive functional quality testing
to examine all operating modes and to guarantee
compliance with the specified technical data. The
testing is performed with testing equipment that
is calibrated to national standards. The statutory
warranty provisions shall be governed by the
laws of the country in which the product was
purchased. In case of any complaints, please
contact your supplier.
Any adjustments, replacements of parts,
mainte-
nance and repair may be carried out only by au
-
thorized technical personnel. Only original parts
may be used for replacing parts relevant to
safety (e.g. power switches, power transformers,
fuses). A safety test must always be performed
after parts relevant to safety have been replaced
(visual inspection, PE conductor test, insulation
resistance measurement, leakage current mea
-
surement, functional test). This helps to ensure
the continued safety of the product.
The
product may only be opened by authorized
and qualified personnel. Prior to working on the
product or before the product is opened, it must
be disconnected from the AC supply network.
Otherwise, personnel will be exposed to the risk
of an electric shock.