The relays operate independently from each other. The terminal port is labeled with C
(common), the other relay ports are labeled with NC (normally closed) and NO (nor-
mally open). Terminal C is connected either with relay port NC or NO.
The type of relays is monostable (or "failsafe"), hence, with no voltage applied, an RF
connection is provided from C to NC, while NO is not connected to C ("Pos = 0"). To
disconnect NC and connect C to NO ("Pos = 1"), a continuous voltage must be applied
to the relay.
The relays are mounted directly into the module's panel and soldered to the module's
printed board, to avoid relay cabling and allow a compact module size.
The switch unit provides power supply and control via a single module bus cable to a
module bus connector on the module's printed board. The module's on-board memory
stores its configuration data.