Functions
R&S
®
OSP
204
User Manual 1178.7100.02 ─ 05
●
Triggering system active and ready for events: both LEDs are green
●
Triggering system active and incoming trigger event:
[Trigger A]-LED is red during approximately 0.5
s, then it returns to green
●
Triggering system active and incoming reset event:
[Trigger B]-LED is red during approximately 1
s, then it returns to green
The
"Trigger Activation"
) provides the following:
●
You can activate or deactivate the trigger functionality.
When triggering is active, this state is indicated by the following:
–
The trigger status LEDs next to the BNC connectors are on (green light).
–
The top right corner of the user interface displays the "TRG" flag:
When the trigger is active, also be aware of the following additional effects:
–
All module views are hidden
–
You cannot switch relays via the
dialog
–
You cannot switch paths via the
dialog, but only by triggering
–
(SCPI), which can interfere with the trigger functionality or can
prevent correct triggering, are not executed.
Calling forbidden commands creates an error message, which you can query
by
.
The following commands are forbidden:
–
Commands for changing path configurations
–
Commands for changing the interconnection configuration
–
Switching commands, which change the state of relays or channels, especially
–
Querying the switching state of relays or channels by
is not
supported
However, the commands
are permitted and deactivate
the trigger mode.
If you cannot activate the trigger, possible reasons are, for example:
–
–
A secondary switch unit does not respond, although it is required for a config-
ured path to be triggered
●
You can send a
"Manual Trigger"
pulse, if the trigger functionality is enabled and
its configuration is complete.
The "Manual Trigger" button is equivalent with sending a remotely controlled trigger
pulse by
or with receiving the next incoming trigger signal at the BNC con-
nector.
The "Manual Trigger" is not available for the
●
You can send a
"Sequence Reset"
for the
trigger, if the trigger func-
tionality is enabled and its configuration is complete.
Configuration