Functions
R&S
®
OSP
207
User Manual 1178.7100.02 ─ 05
When you activate the trigger, the next incoming trigger event switches "Path A". Each
following trigger event toggles from the currently switched path to the other path.
Note:
After the 2
nd
trigger event, your relays are not necessarily in the same state as
before the 1
st
trigger event. For example, consider a relay that is originally
deactiva-
ted
. If "Path A" activates it, but "Path B" does not include this relay, it is therefore not
deactivated. Hence, the relay remains
activated
after the 2
nd
trigger event, before
"Path A" is switched again.
Remote command:
Sequenced ← Trigger Type
Allows triggering 3 to 16 paths in a given sequence.
For configuring the sequenced trigger, first select a "Seq. Length", where the sequence
length
n
=
3
to
16. The dialog displays n path selection boxes, labeled "Path 00" for the
1
st
path up to "Path (n-1)" for the n
th
path.
Select a path in each path selection box.
Activation:
When you activate the trigger, the next trigger event ("Manual Trigger" or incoming trig-
ger event at the [TRIGGER A] BNC connector) switches the 1
st
path ("Path 00"). Each
following trigger event switches the next path in the sequential order up to the n
th
path
("Path (n-1)"). The next trigger event recommences with the 1
st
path.
At any time, you can reset the trigger sequence to the 1
st
path. To do so, click
"Sequence Reset" or send a reset event to the[TRIGGER B] BNC connector.
Note:
If a reset is the first trigger event for the sequenced trigger, the following trigger
event has no effect (it does not switch any path), but it is counted.
Note:
After the last trigger event in the sequence, your relays are not necessarily in the
same state as before the 1
st
trigger event. For example, consider a relay that is origi-
nally
deactivated
. If the 1
st
path activates it, but all other paths do not include this
relay, it is therefore not deactivated. Hence, the relay remains
activated
after the last
trigger event, before the 1
st
path is switched again.
Remote command:
TRIGger:TYPE
TRIGger:SEQuence:DEFine:LENGth
TRIGger:SEQuence:DEFine[:ENTRy]
TRIGger:SEQuence:DEFine:LENGth
Addressed ← Trigger Type
Functional
only
in the switch unit
R&S
OSP320
, but without any paths that use a
OSP320 has a rear trigger input D-Sub
9 connector
(labeled 3 in
) with 4 digital input lines on pin
1 to pin
4. With the resulting
2
4
=
16 addresses, you can use this trigger type to address up to 16 paths directly.
For configuring the addressed trigger, first select an "Address Count", where the
address count
n
=
3
to
16. The dialog displays n path selection boxes, labeled "Path
00" for the 1
st
path up to "Path (n-1)" for the n
th
path.
Configuration